3-24
Settings and Measurement
Lower - Level - Upper ......... Specifies the trigger/alarm level(s) for the conditions set in Mode.
• H, L: Input a numeric value.
• Win In, Win Out: Input a numeric value for the upper and lower limit.
Combination (for trigger only)....Sets the combination of trigger conditions set for each channel.
OR: Starts (stops) capturing data when at least one trigger condition is
met.
AND: Starts (stops) capturing data only when all trigger conditions are
met.
Pulse
Specifies the trigger/alarm for pulse input signals.
These conditions can be set when they have been enabled in the AMP settings.
Mode ................................... Specifies the same conditions as for analog CH conditions (see on
page 3-23).
Lower - Level - Upper ......... Specifies the trigger/alarm level(s) for the conditions set in Mode.
• H, L: Input a numeric value