Service
Mini-Convectron
®
Module with DeviceNet
®
Digital Interface
41
6. If any of the resistance values in Table 6-1 are not within specification,
the gauge tube should be replaced.
7. Phone customer service at
1–303–652–4400
or
1–800–776–6543
to
order a new gauge tube.
8. Hold the PC board/gauge tube assembly and
carefully
push the end of
the replacement gauge tube into the tube socket until it stops.
Make sure the PC boards are properly mounted in the front and rear
cover slots.
9. Install the side, front and rear cover plates on the PC board/gauge tube
assembly.
10. Install the eight end plate screws.
11. Connect the Trip Point and DeviceNet cable connectors.
Gauge tube failures are most frequently caused by one of the following
conditions:
• Chemical Etching – exposure to fluorine, chlorine or mercury causes
the sensor wire to become chemically etched and damaged.
• Overpower – if the gauge electronics and gauge tube are connected
while the gauge electronics are powered ON and the gauge tube is at
vacuum, a spike in the bridge voltage can occur at initial contact and
damage the sensor wire. Always turn power OFF or unplug the gauge
tube before plugging the gauge into the gauge electronics.
• Shock – extreme shock, such as dropping a gauge, can cause a weld
to fail.
Содержание Mini-Convectron 275 Series
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Mini Convectron Module with DeviceNet Digital Interface...
Страница 22: ...14 Mini Convectron Module with DeviceNet Digital Interface...
Страница 44: ...36 Mini Convectron Module with DeviceNet Digital Interface...
Страница 46: ...38 Mini Convectron Module with DeviceNet Digital Interface Chapter 6 Figure 6 1 Gauge Tube Removal from Module...
Страница 51: ......