24
Mini-Convectron
®
Module with DeviceNet
®
Digital Interface
Chapter 4
Atmospheric
Adjustment
1. Allow the chamber pressure to rise to atmospheric pressure.
2. Use Table 4-10 to determine the atmospheric pressure (Torr) for your
altitude and enter the data value for the Atmospheric Adjust Service.
Using Input Polled
I/O
When a master polls the module for measured pressure, the format of the
returned pressure value depends on the data type.
To configure the data format for input polled I/O, see Table 4-5.
Table 4-10
Atmospheric Adjustment Parameters
Altitude Above Sea Level (feet)
Air or N
2
Pressure (Torr)
0
760
1,000
733
2,000
707
3,000
681
4,000
656
5,000
632
6,000
609
7,000
586
8,000
564
9,000
543
10,000
523
Table 4-11
760 Torr Example
Parameter
Service
Class
Instance
Data
Gain Adjust
4C
hex
31
hex
1
00 00 3E 44
hex
Table 4-12
Input polled I/O for pressure values
Pressure values are transmitted in low byte to high byte order.
Instance
Typical device data
Data type
Description
1
23 79
hex
UINT
UINT vacuum pressure (760 Torr)
2
00
23 79
hex
STRUCT
BYTE exception status
UINT vacuum pressure
4
00 00 3E 44
hex
REAL
REAL vacuum pressure
5
(default)
00
00 00 3E 44
hex
STRUCT
BYTE exception status
REAL vacuum pressure (760 Torr)
Содержание Mini-Convectron 275 Series
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Mini Convectron Module with DeviceNet Digital Interface...
Страница 22: ...14 Mini Convectron Module with DeviceNet Digital Interface...
Страница 44: ...36 Mini Convectron Module with DeviceNet Digital Interface...
Страница 46: ...38 Mini Convectron Module with DeviceNet Digital Interface Chapter 6 Figure 6 1 Gauge Tube Removal from Module...
Страница 51: ......