background image

 

 

 

 

 

 

Refrigerated immersion 
Coolers 

 

 

C1G, C1GL 
C2G, C2GL 

 

Operating Manual 

 

 

 
 
 

Содержание C1G

Страница 1: ...Refrigerated immersion Coolers C1G C1GL C2G C2GL Operating Manual ...

Страница 2: ......

Страница 3: ... 4 2 Getting started 5 2 1 Unpacking 5 2 2 Installation 5 2 2 1 C1G C2G 6 2 3 Specifications 6 3 Maintenance and service 7 3 1 C1G C2G 7 3 2 Cleaning 8 3 3 Replacement of fuses 8 4 Guarantee 8 5 Service 8 6 Compliance 9 ...

Страница 4: ...y which provides a safety earth ground terminal Before moving disconnect at the power supply socket Do not remove the IEC connector When lifting use top flange by the ventilation grilles at the front and rear or lift from under the base Ensure that the mains switch is easily accessible during use Do not block or restrict ventilation slots If liquid is spilt inside the unit disconnect it from the p...

Страница 5: ...temperature control and to prevent the formation of ice Do not switch on if the temperature of the liquid in the bath is above 100 C the cooler has been tilted by more than 25 during the past six hours the interval since switching off the cooling system is less than 10 minutes Condensed water vapour from the air will dilute water glycol mixture and eventually ice will form on the cooling coil This...

Страница 6: ...20W Overall power consumption 300VA 500VA Refrigeration gas R134A 100g 160g Dimensions d w h 460 305 225mm 460 305 225mm Flexible pipe length 925mm 925mm Coil Diameter length 77 55mm 77 105mm Electrical Supply 240V 50 Hz 120V 60Hz 240V 50Hz 120V 60Hz Weight 16Kg 21Kg ...

Страница 7: ... removed safely by using refrigerant recovery equipment or return the unit to us for disposal 3 2 Cleaning The cases can be cleaned with a damp cloth after disconnection Do not use solvents The immersed parts can be cleaned with soapy water Before using any decontamination or cleaning method except that recommended check with our Service Department or in other countries with our distributor that t...

Страница 8: ...rvice repairs outside of the guarantee period carried out by Grant Instruments carry a further one year guarantee 4 Service For service return for repair to our Service Department in the UK or in other countries to our distributor Grant Instruments Cambridge Ltd Service Department SHEPRETH Cambridgeshire SG8 6GB England Tel 44 0 1763 260811 E mail service grantinstruments com ...

Страница 9: ...048TZ For other countries please contact your equipment supplier For General WEEE information please visit www b2bcompliance org uk 6 2 RoHS Directive All the products covered by this manual comply with the requirements of the RoHS Directive Directive 2011 65 EC This means the products are free of Lead and other hazardous substances covered by the directive 6 3 Electrical safety and Electromagneti...

Страница 10: ...Notes ...

Страница 11: ......

Страница 12: ...hepreth Grant Instruments Europe B V Cambridgeshire Strawinskylaan 411 SG8 6GB WTC Tower A 4th Floor UK 1077 XX AMSTERDAM THE NETHERLANDS Tel 44 0 1763 260811 salesdesk grantinstruments com grant eu grantinstruments com www grantinstruments com 16601 V9 DMN S90 July 2021 ...