2. Getting started
2.1
Unit overview
2.2
Unpacking
Optima low temperature circulators provide a source of cooling for many sensitive
analytical procedures. Applications requiring the precision temperature control of
samples include spectrophotometry, viscometry, refractrometry and electrophoresis.
Grant’s range of more powerful chillers are ideally suited for applications requiring the
removal of the mechanical or electrical heat produced in apparatus or machinery,
combined with a higher pressure flow for faster heat removal within restricted systems.
Remove packing materials carefully, and retain for future shipment or storage of the unit.
Packs should contain:
2.3
Installation
To avoid condensation on internal components, allow the unit to reach room temperature
before switching it on.
Let the unit stand in its intended working position for 6 hours before switching on. This is
to allow the oil to drain to the bottom of the compressor.
Position the equipment so that there is free air flow through it. Allow at least 100mm
clearance from obstructions on all sides.
COOLER UNIT
Lid
Mains cable
Communications cable
Interconnecting power cable
Operating instructions
Operating voltage
R1
R1
R2
R3
R4
R5
R1L R2L R3L R4L
240V 240V 240V 240V
240V
115V 115V
115V
115V
Page 6
Low temperature bath/circulator
Operating instructions
Version 4 - Jan 2010