Grant-bio Platform Rocker PMR-100 Скачать руководство пользователя страница 12

Platform Rocker

/

PMR-100

/

1.02

Grant Instruments
(Cambridge) Ltd

Shepreth
Cambridgeshire
SG8 6GB
UK
Tel: +44 (0) 1763 260811
Fax: +44 (0) 1763 262410
Email: [email protected]
www.grantinstruments.com

Содержание Platform Rocker PMR-100

Страница 1: ...Operating instructions Platform Rocker PMR 100 For version V 1GW ...

Страница 2: ...Page 2 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Страница 3: ...3 Contents 1 Safety 4 2 General Information 5 3 Getting Started 6 4 Operation of PMR 100 7 5 Specifications 9 6 Guarantee and service 10 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Страница 4: ... do not use extension lead without grounding Only use the external power supply unit provided with this product Ensure that the external power supply connector is easily accessible during use Before moving disconnect from the mains To turn off the unit disconnect the external power supply from the mains outlet If liquid is spilled inside the unit disconnect it from the mains and have it checked by...

Страница 5: ...protein electrophoresis Optimal for biomolecular hybridization on strips and for staining distaining procedures The PMR 100 can be installed inside an incubator ideal for cell growth in single use plastic bags working volumes of bags 5 litres Platform Rocker PMR 100 is equipped with a system of platform overload automatic detection In the case of overload an error message is shown on the unit s di...

Страница 6: ...n of conformity 1 copy 3 3 Set up place the unit on an even firm stable horizontal surface connect the power cord to the external power supply plug the external power supply unit into the 12 V socket at the rear side of the unit 3 4 Platform installation install the platform on the unit and remove the rubber pad secure the platform on the supporting platform on top of the unit with the four suppli...

Страница 7: ...of increment becomes bigger Note If the operation time is not set and the timer indicator shows 0 00 pressing RUN key Fig 1 z will allow continuous use until the STOP key Fig 1 is pressed again 4 4 Speed set Speed RPM With the help of the p and q SPEED keys Fig 1 x set the required rocking intensity in revolutions per minute increment 1 oscill min If the button is pressed for longer time the speed...

Страница 8: ...ws the flashing reading OVERLOAD accompanied by the repetitive sound signal until the STOP key Fig 1 is pressed Eliminate the causes of overload before restarting operation see p 4 6 4 9 After the set time elapses the platform motion stops and the timer shows the flashing reading STOPaccompanied by the repetitive sound signal until the STOPkey is pressed 4 10 If necessary the unit can be stopped a...

Страница 9: ...nge 10 Digital time setting 1 min 99 hrs 59 min non stop Time setting resolution 1 min Maximum continuous operation time 168 hrs Maximum loading 5 kg Platform working area 480 x 380 mm Dimensions 400x480x250 mm Input current power consumption 12V 1 1 A 13W External power supply input AC 100 240 V 50 60Hz output DC 12V Weight not more 13 kg Grant is committed to a continuous programme of improvemen...

Страница 10: ...below return to our service department in the UK or in other countries our distributor 6 3 Cleaning Disinfection Cleaning liquids that do not contain concentrated organic solvents alkali or acid can be used for cleaning the device Standard ethanol 75 can be used for disinfection Before using any other cleaning or decontamination method check with the manufacturer or supplier to make sure that the ...

Страница 11: ...Page 11 Platform Rocker PMR 100 Operating instructions Version 1 02 April 2013 ...

Страница 12: ...atform Rocker PMR 100 1 02 Grant Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB UK Tel 44 0 1763 260811 Fax 44 0 1763 262410 Email scientificsales grantinstruments com www grantinstruments com ...