5.
Once you enter the weight value, double-press on the
key to validate
and move to next step. The display will show the blinking message
“
“ while acquiring the
calibration value.
6.
Lastly, it will show the message
“
”
for a few seconds, asking for the
code of the geographical location where you did the calibration. The
geographical location code is a value from 0 to 31, which you have to
choose from the attached table. Use the
and
→
keys to change the
value and validate by pressing on the
key.
7.
Lastly, the message “
”
will briefly appear, indicating that the
calibration has been saved in the non-volatile memory. The indicator
returns to normal use mode, displaying the weight on the load receptor.
The geographical area where the scale is used can be modified later
whenever you wish by entering the menu with
→
→
→
→
→
(for nn {0-31}).
The automatic correction of the scale calibration according to geographical
area can be disabled by entering the menu with
→
→
→
→
→
.