14.3.
Scale calibration
For the calibration of the scale, it is recommended to use a standard weight.
Calibration procedure using a known weight (automatically
sets the initial zero and the slope value).
Gravity adjustment depending on the
scale’s
geographical
location:
ON / OFF correction (activates/deactivates automatic
correction according to geographical location).
Geographical location code (see attached table).
Manual input (keypad) of the initial zero (A/DC counts value).
Manual input of the span slope, 5 digits. The divisor needed
to transform A/D converter counts to the measurement unit.
Prints a ticket with the configuration and calibration settings
in the device’s memory.
14.4.
Scale calibration procedure
1.
With the platform free of any load, select the
“
” option
.
2.
The display will show that the initial zero value is in progress with the
blinking message
“
”
.
3.
Once the zero value has been calibrated, place the calibration weight (a
known mass weight) on the load receptor.
4.
Enter the weight value in the indicator, including the decimal positions.
Use the cursor movement keys to move through the different positions
on the display.