Страница 1: ...SERIES FH 100 200 2000 6000 OPERATION MANUAL...
Страница 2: ...nd weighing pan 7 Connect the Power Adapter 7 Install the External Facilities 7 Warm up Time 7 Adjust the Level Gauge Purpose 8 Display and keys description 8 Calibration 10 Procedure 10 Basic weighin...
Страница 3: ...n description 20 Routine maintenance and troubleshooting 20 Troubleshooting 20 Clearance 21 Clearing the stainless steel surface 21 Safety check 22 Troubleshooting 22 Technical conditions 24 Guarantee...
Страница 4: ...nstructions Excessive electromagnetic interference will make the displayed value in this balance change Once the interference is dismissed the balance could operate in normal way Please avoid any liqu...
Страница 5: ...ion Please do not open the balance If the guarantee lable is damaged our quality guarantee will automatically cease being effective If the balance does not work well please contact your local distribu...
Страница 6: ...LANCE Put it in a proper place please avoid the following situations Much heat and direct sunlight Air currents and vibrations Excessive moisture Usage Conditions Please do not lay the balance in a qu...
Страница 7: ...dapter Only the power adapter offered by us is allowed Connect the power adapter to the balance Connect the power adapter to the power socket Install the External Facilities Turn off the balance befor...
Страница 8: ...s are used to adjust the level guage Spin the two bolts as shown in the figure until the bubble in the level gauge is in the ceter of the circle Normally it needs to be adjusted repeatedly L R R R L L...
Страница 9: ...shown is given in percent weight PCS Reading shown is given in as a counting The balance is developing a stable reading UNABLE Error operation HHHHH The weight on the pan excedes the capacity of the b...
Страница 10: ...e stable Procedure 1 Adjust the level guage and warm it up for 25 minutes 2 Press the key 3 Press the key the display will read its full range such as 3000 4 Press the key repeatedly it reads the cali...
Страница 11: ...PCS Procedure 1 Place the empty container on the weighing pan 2 Press the key 3 Choose the amount of the reference samples Press the key the balance reads qTy10 press the key choose the number of sam...
Страница 12: ...0 which relates to the accuracy 100 00 of the balance Meanwhile it displays 3 Remove the reference object the balance Reads 0 000 or 0 00 4 Place the goal object on the weighing pan 5 Wait for the OK...
Страница 13: ...be printed out WEIGHING UNITS 1 Press the key to select the desired unit weight g oz dwt or weighing mode normal weighing counting or percentage mode Enabling weighing units The units function can be...
Страница 14: ...14...
Страница 15: ...15...
Страница 16: ...NSTANT Print print immediately after press the key INTERVAL Print print at predeterminated time intervals Note The print function is separate from the line feed setup i e set the print function first...
Страница 17: ...the key while pressing the key the display will show PrInT 2 Press the key the balance reads bAud 3 Press the key the balance reads 300 4 Press the key repeatedly the balance display the other baud r...
Страница 18: ...TER The balance keyboard functions can be accessed via the RS 232 interface The following commands are available U units conversion T tare function C calibration with external standard weight P print...
Страница 19: ...our balance will transmit G for grams O for ounces and C for carats K Stable character it corresponds to the OK indicator on the display S means the reading is stable space means the reading is not st...
Страница 20: ...er 15 characters It is suggested that the m ximum recommended cable lenght is 15 meters the load impedance of the device connected should be between 3000 and 7000 ohms with no more than 2500 pf shunt...
Страница 21: ...All the stainless Steel components need to clean regularly Remove the weighing pan and thoroughl clean it with a damp cloth or sponge The cleaner applicable to stainless Steel are recommended Wipe the...
Страница 22: ...training TROUBLESHOOTING DISPLAY REASON SOLUTION The balance could not get a stable reading or the transducer is damaged Contact your local service center HHHHH The real weight is 5 higher than its c...
Страница 23: ...vibration or air current or there is wrong connect with the pan Put the balance in another place Clear the wrong connect NO DISPLAY No work voltage no connection with transformer Check the power supp...
Страница 24: ...24 TECHNICAL CONDITIONS AC DC adapter Input 220V 110V output 7 5V 15 to 20 48 60Hz Power consumption Maximam 16W Average 8W...
Страница 25: ...fect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary for the reparation of the scale This guarantee does not cover the failures caused by an inap...