Setup F.S.

[FULL  ]

Input F.S.:Press 

 

to choose 

the digit bit; press 

 

to select 

the digit; press 

 

to confirm 

the input F.S. e.g.: 30000

4

Zero point calibration:
Press 

 

when the stable 

light is on

[noLoAd] Assure there’s no loading

5

Load standard weights.

6

F.S. calibration: (Press 

 

directly if the value to be 
input is same as F.S. when 
stable light is on)

[AdLoAd]

Input weight of standard weights 
loaded: Press 

 

to choose 

the digit bit; press 

 

to select 

the digit. Until the input value is 
same as the loaded weight and 
the digit bit is at the highest bit, 
press 

 

to confirm when the 

stable light is on.

7

Calibration finished.

ERROR INDICATION

1

Err  1

AD value is too small when calibration.

2

Err  2

Zero point is out of range when calibration.

3

Err  3

Zero point is out of range when power on

4

Err  5

Weight input is 0 when full scale calibration

5

bAt-Lo

Indicator out of battery, please charge!

6

Err  8

Mistakes in inputting F.S.

7

Err  10

Resolving capability is too high

Содержание C6-10T

Страница 1: ...SERIES C6 3T 5T 10T 15T 20T OPERATION MANUAL ...

Страница 2: ... Front and back view of the indicator 4 Keys function 4 Operation 5 Turning on 5 Zero function 5 Tare 5 Manual accumulation 5 Check accumulation 6 Clear accumulation 6 Parameters setup 6 Calibration 8 Error indication 9 Rechargeable lithium battery 10 Maintenance 10 Guarantee 12 ...

Страница 3: ... times second Display Display 0 999999 5 digits LCD LED 7 indicators Division Value 1 2 5 10 20 50 optional Ambient Environment Distance between scale and indicator 150m Power Supply DC 3 7V Ambient Temperature 0 40 ºC Storing Temperature 25 ºC 55 ºC Ambient Moisture 85 RH Warm up Time 10 15 分钟 ...

Страница 4: ... than 5s it enters setup mode All the setting in this mode should be recorded with the calibration header on or the data will lose Tare function in weighing status Zero function in weighing status For selecting Weighing accumulation Long press for 1sec to turn on Long Press for 3sec to turn off ...

Страница 5: ... open both indicator and scale change the number of transcoder inside with same number and restart Tare At weighing mode press to deduct displaying weight while it is positive and stabilization light is on At no load status of the scale press to deduct tare and then tare light is off Manual accumulation At weighing mode press it displays ADD 01 01 means times of accumulation the Max times is 99 an...

Страница 6: ...he current accumulation PARAMETERS SETUP In weighing status hold for more than 5s and entering setup mode P mode all the setting in this mode should be recorded with the calibration header on or the data will lose P1 P13 parameters are available press to switch and press to select parameter Detail for parameter 1 P1 x kg Lb shift X 1 kg mode default X 2 Lb mode 2 P2 x Not available for this model ...

Страница 7: ...e X 1 No power saving X 2 Power saving available X 3 None 7 P7 x Zero tracking Scope X 1 0 5e X 2 1 0e X 3 1 5e default X 4 2 0e X 5 2 5e X 6 3 0e X 7 3 5e 8 P8 x Zero key Scope X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS 9 P9 x Zero scope upon start X 1 2 FS X 2 4 FS X 3 10 FS default X 4 20 FS ...

Страница 8: ...lue Press to select division value then press to confirm CALIBRATION Turn on the crane scale first then hold key and power on the indicator secondly it enters calibration mode Step Operation Display Explanation 1 Press to choose division value several times d Division value 1 2 5 10 20 50 optional press to confirm e g 2 2 Press to choose decimal point position several times P Decimal point optiona...

Страница 9: ...weights loaded Press to choose the digit bit press to select the digit Until the input value is same as the loaded weight and the digit bit is at the highest bit press to confirm when the stable light is on 7 Calibration finished ERROR INDICATION 1 Err 1 AD value is too small when calibration 2 Err 2 Zero point is out of range when calibration 3 Err 3 Zero point is out of range when power on 4 Err...

Страница 10: ...t be placed directly under sunshine and should be set in the plain space 2 The indicator can t be placed into the place where the dust pollution and vibration are serious 3 Load cell should connect with indicator reliably and the system should be well connected into ground The indicator must be protected from high electrical field and high magnetic field In order to protect the operator indicator ...

Страница 11: ...ction may happen 8 The storage is not granted the free repair guarantee because it is easily exhausted products In order to prolong using life please charge the cell fully before using it If you don t use the indicator for a long time you must charge the cell every two month and for eight hours each charging time Moving or installation must be carefully taken and must avoid strong vibration impact...

Страница 12: ...ation defect of the material During this period GRAM PRECISION SL covers the manpower and the spare parts necessary to repair the scale This guarantee does not cover the failures caused by an inappropriate use or overload The guarantee does not cover the freight cost transport necessary to repair the balance ...

Страница 13: ...13 NOTES ...

Страница 14: ...14 NOTES ...

Страница 15: ...15 NOTES ...

Страница 16: ...Gram Precision S L Travesía Industrial 11 08907 Hospitalet de Llobregat Barcelona Spain Tel 34 902 208 000 34 93 300 33 32 Fax 34 93 300 66 98 comercial gram es www gram group com 003 19072018 ...

Отзывы: