Set System Parameters
3A0294J
61
9.
Remove any excess material from the dispense
valve, if necessary. Include this material in the
weight of the shot.
10. Weigh the shot on the scale.
11. Press
and
to enter the shot
weight in the weight field.
NOTE: The weight must be entered in GRAMS
.
12. Set the scale to zero.
13. Perform steps 8-12 three more times.
After you have entered the fourth shot weight in the
weight field, the material delta top and bottom fields (cal-
ibration values) will automatically update and the spe-
cific weight
on setup screen 1 will update. The
calibration procedure is now complete.
Set Shot Accuracy Errors
Use this screen to set the shot volume deviation from
the shot setpoint that will cause an advisory or alarm.
NOTE:
If either the advisory field or alarm field is set to
zero, the error is disabled.
1.
Navigate to setup screen 11.
2.
Press
to access fields to make changes.
3.
Press
to navigate to the advisory field.
4.
Press
to enter the advisory field.
5.
Press
and
to set the value in the
advisory field.
6.
Press
to enter the value.
7.
Follow steps 4 through 7 for the alarm field.
8.
Press
to exit edit mode.
Содержание SmartWare 262370
Страница 77: ...Parts 3A0294J 77 Parts Kits 262370 and 262372 30 33 28 29 29 28 34 5 2 29 28 4 31 28 29 32 3b 36 38 41 37 3a 3c...
Страница 80: ...Parts 80 3A0294J Kit 262371 25 4 5 28 24 23 27 23 24 26 24 23 2 3c 24 23 3a 3b 30 32 33 31...
Страница 81: ...Parts 3A0294J 81 Kit 262371 continued 17 20 21 18 19 35 47 44 37 37 37 42 38 40 46 39 43 36 45 48 14...
Страница 84: ...Parts 84 3A0294J Kit 262373 continued 39 40 41 42 42 43 44 45 46 47 49 42 50 47 52 53 15...
Страница 86: ...Parts 86 3A0294J Kit 262374 19 20 16 17 18 22 23 4 3a 5 2 24 25 27 29 28 32 3b 3c...
Страница 87: ...Parts 3A0294J 87 Kit 262374 continued 33 34 35 36 36 37 38 39 40 41 43 36 44 43 46 47 13...
Страница 92: ...Parts 92 3A0294J...
Страница 107: ...Technical Data 3A0294J 107...