Preset Mode
14
3A6673B
3.
Use the UP or DOWN ARROW button on
the keypad to highlight CALIBRATE (F
14) and then press the center ENTER
button on the keypad to select the
CALIBRATE option.
4.
The calibration k-factor screen shown in
F
. 15 displays.
NOTE:
The number shown on the
display is the k-factor the meter is
currently using. After completing the
calibration procedure this number may
be the same or different than what is
currently shown on the display.
5.
Select ACTIVATE and press the center
ENTER button on the keypad to begin
metered dispense valve calibration (F
NOTE:
The +/- on this screen is used for
Manual Calibration. See
6.
Dispense exactly one quart or one liter of
fluid into the one quart or one liter clean,
calibrated, volumetric flask.
NOTE:
During the dispense the metered
dispense valve will not display the
volume dispensed. The volume
dispensed is only determined by the
flask measurement. A screen displays
(F
. 16) during the calibration dispense.
7.
When
exactly
one quart or one liter of
fluid is dispensed into the flask use the
center ENTER button on the keypad to
select END. A screen displaying the
quantity of fluid dispensed displays (F
17 shows an example of the dispensed
volume screen).