Troubleshooting
3A5411J
13
Monitor readings are inaccu-
rate.
Tank geometry incorrectly
defined.
See
Tank Level Monitor Soft-
ware Guide
or
PC Software
Guide
for details.
Tank Level Monitor has not
been updated with latest
adjustments made within The
profile ID on the Pulse sys-
tem.
Manually push DISPLAY but-
ton on TLM.
Fluid surface is moving while
reading is being taken.
Ensure pump is not operating
and that nothing is disturbing
the surface of the fluid during
readings.
Pipe adapters installed in the
tank bung.
Ensure there are no adapters
installed in the tank bung.
Tank Level Monitor is not per-
pendicular to top of fluid.
Level tank so that it is per-
pendicular to top of the fluid
and/or realign Tank Level
Monitor.
Inside tank obstruction.
Install in different tank bung
to avoid tank obstruction.
Device is unable to register
to the Pulse Pro network or
the device does not rejoin
the network after system
reboot.
Device is not meshing prop-
erly with the device network.
Toggle to Discovery mode on
the HUB. Wait one minute
and toggle it back. Press the
button to attempt a network
connection. If this does not
resolve the issue, remove the
batteries, wait two minutes
then reinstall the batteries.
Problem
Cause
Solution