Aquamax KF
Portable
v.13.08
28
PRINT-OUTS
The Aquamax KF
Portable
allows you to select different print modes. If you choose to have the printer on,
you will receive a hard copy of the result obtained, together with the selected parameters for that particular
titration every time a titration is performed. Options available;
Off
- no result print out
Every result
- will print result after every titration, first result also prints all calculation input data.
Full Report
- will print results after every titration and then print out statistical data after a selected
number of titrations.
The Aquamax KF
Portable
automatically increments the next run number every time you press the start
button. The run counter is reset to zero when power is switched off and back on again. A duplicate set of
results can be printed out simply by holding down the start button for 3 seconds. Results memory is cleared
after duplicate print out or when power is switched off.
PRINT OUT EXPLANATION
EXAMPLE OF V/SG CALCULATION
Titrator model
Software version
Titrator serial number
Date titrator calibrated
Date sample run
Run number & time of analysis (F = generator electrode with frit)
Nett microgram count (after drift correction)
Drift value at start of titration
Result format selected
Calculation mode selected
Sample volume
Sample density
Calculated result water content
Subsequent runs of the same sample and calculation parameters show run
number, run time, nett count,
Drift value and calculated result.
Run number & time of analysis
EXAMPLE OF W/w CALCULATION
Run number & time of analysis (N = generator electrode without frit)
Nett microgram count (after drift correction)
Drift value at start of titration
Result format selected
Calculation mode selected
Weight of syringe plus sample
Weight of syringe after sample injected
Weight of sample analysed
Calculated result water content
Aquamax KF Portable
V xxxx
Serial No.
71000
Calibrated DD/MM/YYYY
Run Date
Run: 1 Run Time
F
Nett Count
0.0 µg
Drift
00
Result Format
%
Calculation Mode V/SG
Volume
1.0000 ml
Density
1.0000
Res:
0.0000%
Run: 2
Run Time
Nett Count
0.0 µg
Drift
00
Res:
0.0000%
Aquamax KF Portable
V xxxx
Serial No.
71000
Calibrated DD/MM/YYYY
Run Date
Run: 1 Run Time
N
Nett Count
0.0 µg
Drift
00
Result Format
%
Calculation Mode W/w
Total Weight
0.0000 gm
Tare Weight
0.0000 gm
Sample Weight 0.0000 gm
Res:
0.0000%