HOW TO RUN A TEST
Please read the section PROGRAMMING A METHOD then use the PROGRAMMING FLOW CHART
before running any samples,
Example: Transformer oil sample by V/SG
Program the Aquamax KF
Portable
with parameters for the analysis. For Transformer Oil samples, these
parameters are usually:-
Result Format
= mg/kg or ppm
Calculation Mode
= V/SG
Sample Volume
= 1.0 ml
Sample SG
= 0.875
Other parameters such as Sample ID number, Printer Mode etc are optional.
ANALYSIS METHOD
1.
Confirm that Aquamax KF
Portable
is in “Ready” mode
2.
Flush 1.0ml syringe several times (minimum 6 times) with sample
3.
Fit luer needle and flush through with sample
4.
Draw sample into syringe beyond the 1.0ml marking
5.
Invert syringe so that any air bubbles can be ejected through the needle and adjust syringe
plunger to the 1.0 ml mark
6.
Wipe off excess sample from outside of needle using a clean, dry tissue or paper towel
7.
Pierce needle through injection septa of titration vessel (1 – 2 cm)
8.
Press START
9.
Push needle into anode reagent and inject sample
10.
Withdraw needle from titration vessel
11.
Read result, in mg/kg (ppm) water, on display and printout
12.
Repeat steps 2 – 11 if duplicate result required