![Gow-Mac 5900 Series Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/gow-mac/5900-series/5900-series_operating-manual_2248318006.webp)
GOW-MAC Instrument Co.
6
B. All gas cylinders in use and in storage
MUST
be secured to an immovable
structure to prevent acciden tal falling or movement. Read the relevant safety
codes.
C.
DO NOT
move or transport cylinders with regulators attached or without safety
cap screwed on over the valve system.
D. Store or move cylinders
ONLY
in the vertical position.
E. Store cylinders in a well ventilated area away from heat or ignition sources.
F. When installing tubing, provide
ONLY
approved, adequate pressure reducing
regulators and pressure relief devices to prevent over-pressurizing the tubing and
equipment.
WARNING
READ CGA AND/OR O.S.H.A PROCEDURES FOR PROPER HANDLING OF
COMPRESSED GAS CYLINDERS.
1.5 GENERAL
A. Perform periodic leak checks at all fitting areas. The GOW-MAC Model 21-070
Mini Gas Leak Detector is recommended to find small leaks.
B. Keep combustibles away from gas cylinders and elimi nate ignition sources.
C.
DO NOT
place papers, charts, samples, etc. on top of the GC.
D. Maintain adequate ventilation. Vent all gases to the outside. DO NOT allow
flammable and/or toxic wastes to accumulate.
E. De-pressurize supply gas piping before working on it.
Содержание 5900 Series
Страница 2: ...GOW MAC Instrument Co 2 ...
Страница 26: ...GOW MAC Instrument Co 26 ...
Страница 36: ...GOW MAC Instrument Co 36 ...
Страница 38: ...GOW MAC Instrument Co 38 ...