background image

4

GMC-I Messtechnik GmbH

3

Performing Measurements

The probe should be calibrated on a regular basis in order 
to assure flawless functioning and measurement accuracy. 
Due to the fact that the nickel iron core used for types 
B

F and Clip 0100S is highly sensitive to impact, special 

attention must be paid during measurement, storage and 
transport: Powerful impacts and vibration must be 
avoided.
Make sure that the secondary circuit is connected to the 
measuring circuit at the measuring or recording instrument 
first. Observe the safety precautions listed in chapter 1, as 
well as the specified allowable output load range for the 
respective clip-on transformer. Observe correct phase 
sequencing for k and l at the measuring instrument for 
power measurements.
Type A and D probes are equipped with a current output 
and are connected to the current measurement inputs (e.g. 
300 mA). Before closing the probe around the current con-
ducting cable, check the secondary circuit, as well as the 
measurement input at the measuring instrument with an 
ohmmeter if necessary: Neither circuit may be interrupted!

Do not enclose the primary conductor with the clip until 
after this test has been performed. The measurement value 
is calculated by multiplying the displayed value by the 
transformation ratio, or the appropriate transformation ratio 
can be selected at the measuring instrument (e.g. 

METRA

HIT

B

ASE

METRA

HIT

T

ECH

 or 

METRA

HIT

T-

COM

). 

Be sure to remove the clip from the primary conductor 
before the connections at the measuring instrument are 
disengaged. 
Probe types B, C, E, F and Clip 0100S convert primary 
current into a proportional output voltage, and must there-
fore be connected to the voltage measurement input (mV 
or V) at the measuring instrument. 

Содержание WZ12A

Страница 1: ...WZ12A F Clip 0100S 3 349 016 02 9 10 18 Operating Instructions E ...

Страница 2: ...onduc tor to 400 V phase to phase in 4 wire 3 phase systems or to 500 V phase to phase in 3 wire 3 phase systems An open secondary circuit at the WZ12A or WZ12D may lead to dangerous secondary voltages and to the destruction of the secondary coil Make absolutely cer tain that the secondary circuit is always connected to the measuring circuit at the measuring or recording instru ment first The seco...

Страница 3: ...www gossenmetrawatt com 2 General Description The WZ12 current transformer allows for the interruption free measurement of alternating current within various measuring ranges from 1 mA to 180 A by enclosing the electrical conductor The WZ12A basic model has been designed for general current measurements in electrical circuits of greater than 15 A Its low price and compact dimensions make it an att...

Страница 4: ... D probes are equipped with a current output and are connected to the current measurement inputs e g 300 mA Before closing the probe around the current con ducting cable check the secondary circuit as well as the measurement input at the measuring instrument with an ohmmeter if necessary Neither circuit may be interrupted Do not enclose the primary conductor with the clip until after this test has...

Страница 5: ...A 1 A 150 A Frequency Range 45 65 400 Hz 45 65 500 Hz 45 65 400 Hz 45 65 500 Hz 30 45 65 500 Hz 30 45 65 500 Hz 45 65 500 Hz Transformation Ratio 1000 1 100 mA A 100 mV A 1 mV mA 1000 1 100 mA A 10 mV A 100 mV A 1 mV mA 1 mV A 1mV A Output Burden Load 5 1 M 1 M 50 10 k 11 13 k 1 M 10 k 10 k Intrinsic Uncertainty under Reference Conditions 3 rdg 1 5 rdg 1 mA 3 rdg 0 2 mA 2 5 rdg 1 mA 2 rdg 10 mA 2 ...

Страница 6: ...g 0 1 A Influence Error Temperature 10 K typ 3 rdg 1 5 rdg 3 rdg 2 5 rdg 2 rdg 2 rdg 3 rdg Typical Phase An gle Error 45 65 Hz fmin fmax not defined 3 10 not defined 3 10 2 10 2 15 not defined 3 10 3 10 Max Overload cont dyn 1s 360 A 900 A 200 A 500 A 300 A 750 A 300 A 750 A 300 A 750 A 30 A 75 A 300 A 750 A Open Circuit Voltage max 15 V 1 max 15 V max 27 V max 27 V 1 max 27 V max 27 V max 27 V Ty...

Страница 7: ...0 V Pollution degree 2 Test Voltage 3 7 kV AC 1 min Ambient Conditions Operating Temperature 10 C 40 C Storage Temperature 20 C 70 C Electromagnetic Compatibility Emission EN 61326 1 Immunity EN 61326 1 Mechanical Design Dimensions 40 mm x 26 mm x 120 mm Jaw Opening 15 mm Housing Material Plastic Novodur Color anthracite Connector Cable Length approx 120 cm Protection IP40 WZ12A F with contact pro...

Страница 8: ... basis Remove contamination and dust from the plastic components and connector cables with a cloth which has been dampened with soapy water Mea sure the resistance at the secondary circuit with an ohm meter on a regular basis If loose contacts become appar ent remove the clip on current transformer from service 6 Repair and Replacement Parts Service Calibration Center and Rental Instrument Service...

Отзывы: