Capacity Measurement
1.
Disconnect supply power from the electrical circuit of the device to be measured, and discharge all high-voltage capacitors
2.
Make sure that the device under test is voltage-free. Capacitors must always be discharged before measurement is performed.
3.
Set the rotary switch to „
“
4.
Connect the (discharged!) device under test to the sockets with the measurement cables as shown
8