Temperature Measurement
Measurement with Thermocouples, Temp TC:
1.
Set the rotary switch to „Temp
TC
”
2.
The reference temperature is measured at the internal reference junction
3.
Connect the sensor to the two accessible jacks. The instrument displays the measured temperature using the selected unit of
measure
Measurement with Resistance Thermometers
Set the rotary switch to „Temp
TC
“ or „Temp
RTD
”
Automatic Compensation:
1.
Press the ZERO | ESC key. „Short leads“ appears on the display
If you prefer to enter cable resistance directly, you can skip the following entry prompt
2.
Short circuit the measuring instrument’s connector cables. “000.00” appears at the display. After pressing the FUNC | ENTER key,
automatic compensation of cable resistance is activated for all subsequent measurements. The short-circuit can now be liminated,
and the device is ready for use
Entering Cable Resistance:
1.
Press the ZERO | ESC key once again in the automatic compensation menu
2.
Enter the known resistance of the connector cables with the scroll keys: Select the digit to be changes with the < > keys, and
change the respectively selected digit with the up, down keys. The default value is 0.43 Ω. Values can be selected within a range
of 0 to 50 Ω
3.
Upon pressing the FUNC | ENTER key, the selected value is activated and the display is returned to the measuring function. Cable
resistance remains in memory even after the instrument has been switched off