Gossen MetraWatt METRALINE RCD CHECK Скачать руководство пользователя страница 4

4

GMC-I Messtechnik GmbH

2.2

Operation and OLED Display

1 Graphic OLED display 
2

START

 key: 

– 

Switch on

Press and hold until the display 
lights up. 
– 

Start measurement

Press and hold until measurement 
starts. Briefly pressing starts mea-
surement if voltage is applied to the 
test probes (exception for Uc: is dis-
played immediately after voltage is 
applied). 
– 

Measurement point illumina-

tion

Press briefly to switch illumination 
on and off. 
– 

Switch off

Press twice briefly to switch the 
instrument off.

3

FUNC

 key for selecting the following 

measuring functions: 

Uc, TIME, 

I

4

I

key for selecting nominal 

residual current

5

TYPE key:

– 

Uc function

RCD Type Selection

– TIME function

Selection of waveform and initial polarity of the residual current, as well as 
RCD type 

– I function

Selection of waveform and initial polarity of the residual current

6

xI

N

 key 

– Uc function

Selection of the touch voltage limit value

– TIME function

: selection of a multiplier for nom. residual current 

Figure 2.3:

Control Panel and OLED Display

Figure 2.4:

Example of Displayed Information

The information which appears at the display varies depending on 
the selected function.

2.3

Initial Start-Up

After inserting the batteries in accordance with section 5.1 on 
page 9, the measuring instrument is ready for operation.

3

Performing Measurements

3.1

Switching the Instrument On and Off

The instrument is switched on by pressing and holding the START 
key. Briefly press the START key twice in order to switch the 
instrument off, during which no voltage may be applied to the test 
probes! The instrument is switched to the standby mode after 
several seconds (reduced brightness), if none of the keys has 
been pressed and no voltage is applied to the test probes. The 
instrument is switched out of the standby mode (i.e. back to full 
brightness) by pressing any key or applying voltage to the test 
probes. The instrument is shutdown automatically if it has 
remained inactive for about 1 minute, i.e. if no keys have been 
pressed and no voltage has been applied to the test probes dur-
ing this time.

3.2

Instructions and Principles with Validity for 
All Measurements

Setting Parameters

The desired functions or parameters are selected with the 

FUNC

I

N

TYPE

 and 

xI

keys (see section 3.3). Measurement is trig-

gered by pressing the 

START

 key. All selected functions or param-

eters remain valid until they are changed.

Interference Voltage

If (interference) voltage of less than 190 V or greater than 255 V is 
applied to the test probes, corresponding information appears at 
the display. If this is the case, the START key is disabled.

Blown Fuse

If line voltage is applied to the test probes, and if the fuse is either 
blown or has not been inserted, this is indicated at the display.

Figure 3.3:

Blown Fuse Display

The fuse must be replaced as described in section 5.1.3.

Note

If a voltage within a range of 25 to 190 V is applied to the 
test probes, the blown fuse symbol is also displayed at 
the instrument even if the fuse is OK. For this reason, 
make sure that voltage in the measured electrical circuit is 
really less than 190 V before replacing the fuse.

Prerequisites for RCD Measurement

If a voltage within a range of 190 to 255 V is applied to the test 
probes, the TRMS value appears at the display and measure-
ment can be triggered by pressing the 

START

 key.

Figure 3.4:

Example of a Voltage Measurement 

For safety reasons, the test instrument first determines 
whether or not touch voltage Uc is higher than the selected 
limit value each time the 

START

 key is pressed before measur-

ing RCD parameters. If the limit value is exceeded, this is indi-
cated at the display and the measurement is ended.

3

1

2

4

5

6

Battery Display

Measuring

I

and

Temperature Display

RCD Type

Ancillary

Main
Reading

Function

I

N

 

Value

Multi-
pliers

Waveform

Figure 3.1:

Display of (Inter-
ference) Voltage 
< 190 V

Figure 3.2:

Display of (Inter-
ference) Voltage 
> 255 V

Содержание METRALINE RCD CHECK

Страница 1: ...Operating Instructions METRALINE RCDCHECK RCD Test Instrument 3 349 694 03 3 6 13 ...

Страница 2: ... electrical voltage Warning concerning a source of danger attention observe documentation EC mark of conformity This instrument fulfills all require ments of applicable European directives Before using the instrument it must be assured that it is safe Do not use if Visible damage is apparent The battery compartment lid is missing The device has been stored under unfavorable conditions for a length...

Страница 3: ...is enclosed in a compact housing with a pat ented means of retaining the test probes The high contrast four color OLED display assures excellent legi bility When performing measurements under unfavorable light conditions measuring point illumination can be switched on white LED at the front The RCD test instrument is capable of the following measure ments RCD disconnection time RCD breaking curren...

Страница 4: ... the test probes The instrument is shutdown automatically if it has remained inactive for about 1 minute i e if no keys have been pressed and no voltage has been applied to the test probes dur ing this time 3 2 Instructions and Principles with Validity for All Measurements Setting Parameters The desired functions or parameters are selected with the FUNC I N TYPE and xI N keys see section 3 3 Measu...

Страница 5: ...bes from the device under test during measurement Measurement results might oth erwise be distorted Influences on Measurement Results Current flowing via PE has an unfavorable influence on mea surement results or makes measurement entirely impossible due to for example consuming devices or capacitance between L and PE Consuming devices of this sort should therefore be removed prior to measurement ...

Страница 6: ...EN 61009 and IEC 60364 4 41 standards An RCD may not trip at I N Select the TIME function by pressing the FUNC key The nominal value for residual current can be set with the I N key RCD type is selected with the TYPE key Select a multiplier with the help of the xI N key Figure 3 15 Example Settings for Time to Trip Measurement Connect the instrument to L and PE downstream from the RCD as shown in ...

Страница 7: ...e to earth must be 190 V at 45 to 65 Hz dis played results are otherwise not reliable Querying the Firmware Version Before querying the firmware version disconnect both test probes from the device under test measuring circuit and switch the test instrument off Press and hold the FUNC key while switching the test instru ment on by pressing and holding the START key The firmware version appears at t...

Страница 8: ...humidity 40 to 60 Device position any Ambient Conditions Operating Conditions Operating temperature 0 to 40 C Relative Humidity max 85 no condensation allowed Device position any Storage Conditions Temperature 10 to 70 C Relative Humidity max 90 at 10 to 40 C max 80 at 40 to 70 C Device position any Power Supply Batteries 4 ea AAA LR03 1 5 V alkaline or 1 2 V NIMH with at least 750 mAh Number of m...

Страница 9: ...harger is con nected to the mains and to the charging socket see figure 2 1 Charging takes approximately 5 hours and 30 minutes integrated safety timer if the batteries have been fully depleted Safety Precautions Do not attempt to recharge alkaline batteries they may leak explode etc The test instrument may be severely damaged or destroyed as a result After initially charging new batteries and aft...

Страница 10: ...ned deviations are used to correct display values during later use of the instrument We would be happy to perform DAkkS or factory calibration for you at our calibration laboratory Further information is available at our website www gossenmetrawatt com Company DAkkS Calibration Center or FAQs General Calibration Questions and Answers Recalibration of your instrument at regular intervals is essenti...

Отзывы: