Gossen MetraWatt METRALINE ISO CHECK Скачать руководство пользователя страница 5

GMC-I Messtechnik GmbH

5

3.3.2

Insulation Resistance Measurement

Switch insulation measurement on by pressing the 

R

ISO

/

U

SPD

 

key.

Figure 3.7:

Setting Test Voltage

Select the required measuring voltage with the help of the 

▲ 

U

TEST

 

and 

▼ 

U

TEST

 

keys. By briefly pressing the respective 

key, voltage is increased or decreased to one of the following 
nominal values: 50, 100, 250, 500 or 1000 V. The selected 
test voltage appears at the display in the 

U

TEST

 field. For indi-

vidualized value settings, press and hold either the 

▲ 

U

TEST

 or 

▼ 

U

TEST

 key until the value can be changed in 1 V steps 

within a range of 50 to 1000 V. The desired value is then se-
lected by once again briefly pressing the respective key. Sev-
eral seconds after one of these two keys has been activated 
for the last time, the test instrument is returned to the mode 
which allows for switching amongst nominal values of 50, 
100, 250, 500, 1000 V only.

Contact the device under test with both test probes.

Figure 3.8:

Connection Example

Note

If an interference voltage of greater than about 10 V is 
present in the measured electrical circuit, its value is dis-
played and the 

START

 key is disabled (see section 3.3.1 

on page 4). Eliminate the source of interference voltage 
before continuing with the measurement.

Trigger measurement by pressing the 

START

 key. Release the 

key as soon as measurement is started. 
Rising measuring voltage is displayed by means of a bar 
graph (this may take several tens of seconds in the case of 
large capacitances). At the same time, the value appears at 
the display in the 

U

IN

 field. The measuring cycle is ended auto-

matically.

Note

If measurement needs to be continued for a longer period 
of time, the 

START

 key has to be pressed and held for the 

entire duration. 
In contrast to this, the automatic cycle can be ended pre-
maturely by briefly pressing the 

START

 key. No results are 

displayed in this case.

Read the measured insulation resistance value.

Note

The device under test may not be disconnected from the 
test instrument as long as the “!” warning symbol is lit up. 
Discharging large capacitances may take several tens of 
seconds!

Note

Whistling or hissing noises may emanate from the test 
instrument during measurement.

3.3.3

Measurements at Varistors (surge protection devices)

For this measuring function, the test instrument generates a rising 
direct voltage within a range of 50 to 1000 V and, at the same 
time, measures current flowing through the device under test. As 
soon as current reaches a level of 1 mA, voltage is no longer 
increased and the voltage value for the so-called milliampere 
point appears at the display. In accordance with the respective 
function which has been selected with the 

FUNC

 key, results for 

overcurrent protection devices are also evaluated automatically 
and listed as shown in the following table.

Note

If the 

DCMAX

 function has been selected, the upper voltage 

limit for automatic results evaluation can be adjusted with the 

U

TEST

 

and 

U

TEST

 keys.

If the 

DCMIN

 function has been selected, the lower voltage 

limit for automatic results evaluation can be adjusted with the 

U

TEST

 

and 

U

TEST

 keys.

For actual measurement or evaluation, it doesn’t matter which of 
these two functions is selected, i.e. either 

DCMAX

 or 

DCMIN

* Table for Surge Protection Device

If the 

SPD LIST

 function is selected (SURGE PROTECTION DEVICE 

TABLE), a specific type and manufacturer/distributor is suggested 
at the display, voltage at the milliampere point is displayed and 
further information may appear as well.
The required type of surge protection device can be selected with 
the 

▲ 

U

TEST

 and 

▼ 

U

TEST

 keys.

Switch to off position!

Figure 3.9:

Sample Results for an 

Insulation Measure-
ment (the DUT is 
being discharged)

Figure 3.10:

Insulation Measure-
ment Results (the DUT 
has been fully 
discharged: U

IN

 = 0 V)

Symbol in the Display and its Meaning

Selected function

DC

Voltage at the milliampere point 
is measured

Voltage at the milliampere point 
is outside of the measuring 
range

DCMAX

DCMIN

Voltage at the milliampere point 
is within the specified range

Voltage at the milliampere point 
is outside of the specified range

SPD LIST *

Voltage at the milliampere point 
is within the specified range for 
the selected type of surge pro-
tection device

Voltage at the milliampere point 
is outside of the specified range 
for the selected type of surge 
protection device

USER

USER

USER

USER

Содержание METRALINE ISO CHECK

Страница 1: ...Operating Instructions METRALINE ISOCHECK Insulation Resistance Tester 3 349 691 03 2 4 13 ...

Страница 2: ...ng The device has been stored under unfavorable conditions for a lengthy period of time The device has been handled impermissibly e g has been dropped from a height of 1 meter or more The test instrument does not function as described in these operating instructions if this is the case we recommend resetting the device as described in section 3 5 on page 7 CAUTION Do not touch electrically conduct...

Страница 3: ...s are simultaneously recessed and protected In order to charge batteries which have been inserted into the instrument the flexible test probe s connector plug must be removed and the slide must be pushed to the left so that the socket at the right is made accessible and the charger can be plugged into it Figure 2 2 Detail View of the Back Panel with Battery Compartment Lid 2 2 Control Panel 1 Grap...

Страница 4: ...nce voltage to appear at the display Do not remove the test probes prematurely during the course of the measurement because this may lead to distorted mea surement results After insulation measurement has been completed any remaining residual voltage is displayed as UIN which may result from cable capacitance Contact to the device under test must be main tained for as long as a capacitive DUT is b...

Страница 5: ...y take several tens of seconds Note Whistling or hissing noises may emanate from the test instrument during measurement 3 3 3 Measurements at Varistors surge protection devices For this measuring function the test instrument generates a rising direct voltage within a range of 50 to 1000 V and at the same time measures current flowing through the device under test As soon as current reaches a level...

Страница 6: ...hich may result from cable capacitance Contact to the device under test must be main tained for as long as the capacitive DUT is being discharged via the test instrument s internal resistor The falling voltage value can be observed directly at the UIN display Do not disconnect the DUT until voltage UIN is less than 25 V Comments Before measuring surge protection devices disconnect them from the in...

Страница 7: ...ast 750 mAh Number of measurements with batteries at 800 mAh approx 1 000 measurements with 500 V test voltage on 500 k Electrical Safety Measuring category with safety cap applied to test probe CAT III 300 V without safety cap applied to test probe CAT II 300 V Pollution degree 2 Protection class II Mechanical Design Display OLED multicolored graphic Protection IP 43 Dimensions Approx 260 x 70 x ...

Страница 8: ...d soapy water for cleaning Do not place the test instrument back into service until its surface is completely dry Attention Do not use cleaning agents which contain benzine or al cohol Prevent liquids from penetrating into the test in strument s interior 5 3 Recalibration The measuring tasks performed with your instrument and the stressing it s subjected to influence aging of its components and ma...

Отзывы: