Gossen MetraWatt METRALINE ISO CHECK Скачать руководство пользователя страница 2

2

GMC-I Messtechnik GmbH

Table of Contents

Page

1

Introduction ...................................................................... 2

1.1

Scope of Delivery ...........................................................................2

1.2

Optional Accessories ......................................................................2

1.3

Safety Precautions .........................................................................2

1.4

Applications ..................................................................................3

1.5

Applicable Standards  .....................................................................3

1.6

Environment ..................................................................................3

2

Device Description  ........................................................... 3

2.1

Housing ........................................................................................3

2.2

Control Panel .................................................................................3

2.3

Initial Start-Up  ...............................................................................3

3

Performing Measurements  .............................................. 4

3.1

Switching the Instrument On and Off, Energy-Saving Mode, 
Automatic Shutdown ......................................................................4

3.2

Instructions and Principles with Validity for 
All Measurements ..........................................................................4

3.3

Measuring Functions ......................................................................4

3.3.1

Voltage Measurement  ....................................................................4

3.3.2

Insulation Resistance Measurement  ................................................5

3.3.3

Measurements at Varistors (surge protection devices)  ......................5

3.4

Further Device Functions ................................................................6

3.5

Device Reset Function ....................................................................7

4

Technical Data  ................................................................. 7

4.1

Measuring Functions ......................................................................7

4.2

General Data  .................................................................................7

5

Maintenance .................................................................... 7

5.1

Device Power Supply  .....................................................................7

5.1.1

Inserting and Replacing the Batteries  ..............................................7

5.1.2

Recharging the Batteries  ................................................................8

5.2

Cleaning .......................................................................................8

5.3

Recalibration .................................................................................8

6

Repair and Replacement Parts Service,
Calibration Center and Rental 
Instrument Service ........................................................... 8

7

Product Support  ............................................................... 8

1

Introduction

1.1

Scope of Delivery

1

Test instrument with mobile test probe

4

Batteries (AAA)

1

Pouch

1

Condensed operating instructions

1

CD ROM with operating instructions in available languages

1

Factory calibration certificate

1.2

Optional Accessories

4 ea. rechargeable AAA NiMH batteries (Z507B)
1 battery charger (Z507A)

1.3

Safety Precautions

Read the operating instructions thoroughly and carefully before 
using your instrument. Follow all instructions contained therein. 
Make sure that the operating instructions are available to all users 
of the instrument.

Meanings of Symbols on the Instrument

This device is equipped with double or reinforced insula-
tion.

Danger of injury due to electrical current, 
warning regarding dangerous electrical voltage

Warning concerning a source of danger 
(attention, observe documentation!)

EC mark of conformity: This instrument fulfills all 
requirements of applicable European directives.

Before using the instrument, it must be assured that it is safe. Do 
not use if:

Visible damage is apparent

The battery compartment lid is missing

The device has been stored under unfavorable conditions for 
a lengthy period of time

The device has been handled impermissibly, e.g. has been 
dropped from a height of 1 meter or more

The test instrument does not function as described in these 
operating instructions (if this is the case, we recommend 
resetting the device as described in section 3.5 on page 7)

CAUTION

Do not touch electrically conductive parts, test probes etc. 
when the device is switched on and voltage might still be con-
ducted by a test probe – DANGER OF INJURY!

Only use test probes which are included with the instrument 
or are available as accessories.

The device must be switched off and no voltage may be 
applied when exchanging accessories.

It is absolutely essential to adhere to all safety precautions, 
regulations and standards when performing measurements.

No keys may be pressed when connecting the instrument to a 
device under test.

The test instrument may not be subjected to the influence of 
aggressive substances, gases, vapors, liquids or dust.

The test instrument may only be used under the conditions 
listed in the technical data in section 5 on page 7.

If the device is moved from a colder to a warmer room, con-
densation may occur, in which case a brief period of acclima-
tization is advisable.

We recommend removing the batteries during lengths periods 
of storage.

Two relatively strong magnets are integrated into the test 
instrument. Avoid close proximity to magnetically sensitive 
objects such as watches, credit cards and the like.

The illustrations in these operating instructions are drawings 
and may therefore deviate from reality.

Use original accessories only.

Maximum permissible voltage between test probe and ground 
is 300 V!

Maximum permissible (externally applied) voltage between the 
test probes is 600 V!

Opening of Equipment / Repair

The equipment may be opened only by authorized service per-
sonnel to ensure the safe and correct operation of the equipment 
and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser-
vice personnel.
In case the equipment was opened by unauthorized personnel, 
no warranty regarding personal safety, measurement accuracy, 
conformity with applicable safety measures or any consequential 
damage is granted by the manufacturer.

!

!

Содержание METRALINE ISO CHECK

Страница 1: ...Operating Instructions METRALINE ISOCHECK Insulation Resistance Tester 3 349 691 03 2 4 13 ...

Страница 2: ...ng The device has been stored under unfavorable conditions for a lengthy period of time The device has been handled impermissibly e g has been dropped from a height of 1 meter or more The test instrument does not function as described in these operating instructions if this is the case we recommend resetting the device as described in section 3 5 on page 7 CAUTION Do not touch electrically conduct...

Страница 3: ...s are simultaneously recessed and protected In order to charge batteries which have been inserted into the instrument the flexible test probe s connector plug must be removed and the slide must be pushed to the left so that the socket at the right is made accessible and the charger can be plugged into it Figure 2 2 Detail View of the Back Panel with Battery Compartment Lid 2 2 Control Panel 1 Grap...

Страница 4: ...nce voltage to appear at the display Do not remove the test probes prematurely during the course of the measurement because this may lead to distorted mea surement results After insulation measurement has been completed any remaining residual voltage is displayed as UIN which may result from cable capacitance Contact to the device under test must be main tained for as long as a capacitive DUT is b...

Страница 5: ...y take several tens of seconds Note Whistling or hissing noises may emanate from the test instrument during measurement 3 3 3 Measurements at Varistors surge protection devices For this measuring function the test instrument generates a rising direct voltage within a range of 50 to 1000 V and at the same time measures current flowing through the device under test As soon as current reaches a level...

Страница 6: ...hich may result from cable capacitance Contact to the device under test must be main tained for as long as the capacitive DUT is being discharged via the test instrument s internal resistor The falling voltage value can be observed directly at the UIN display Do not disconnect the DUT until voltage UIN is less than 25 V Comments Before measuring surge protection devices disconnect them from the in...

Страница 7: ...ast 750 mAh Number of measurements with batteries at 800 mAh approx 1 000 measurements with 500 V test voltage on 500 k Electrical Safety Measuring category with safety cap applied to test probe CAT III 300 V without safety cap applied to test probe CAT II 300 V Pollution degree 2 Protection class II Mechanical Design Display OLED multicolored graphic Protection IP 43 Dimensions Approx 260 x 70 x ...

Страница 8: ...d soapy water for cleaning Do not place the test instrument back into service until its surface is completely dry Attention Do not use cleaning agents which contain benzine or al cohol Prevent liquids from penetrating into the test in strument s interior 5 3 Recalibration The measuring tasks performed with your instrument and the stressing it s subjected to influence aging of its components and ma...

Отзывы: