![Gossen MetraWatt METRAHIT AM BASE Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/gossen-metrawatt/metrahit-am-base/metrahit-am-base_operating-instructions-manual_2247247052.webp)
GMC-I Messtechnik GmbH
59
Technical Data
Display
LCD panel (65 mm x 36 mm) with analog and digital display
including unit of measure, type of current and various special
functions
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Analog
Display
LCD scale with bar graph or pointer,
depending upon
A
.
diSP
parameter setting
Scaling
With 4 division lines each
1 bar/pointer corresponds to 500 digits at
the digital display
Polarity display
With automatic switching
Overflow display
With the symbol
Measuring rate
40 per second and display refresh
(U and I)
Digital
Display / Char. Height 7-segment characters / 15 mm
Number of places
4½ place
11,999 steps
Overflow display
“
0L
” is displayed for
12,000 digits
Polarity display
“–” (minus sign) is displayed
if plus pole is connected to “
”
Measuring rate
10 measurements per second;
40 per second with Min/Max function
except with capacitance, frequency and
duty cycle measuring functions
Refresh Rate
2 times per sec., every 500 ms
Acoustic Signals
For voltage
Intermittent signal at above 1000 V
For current
Intermittent signal at above 10 A
Continuous signal at above 16 A
Fuse for METRAHIT AM X-TRA / AM TECH / AM PRO
METRAHIT X-
TRA
/ O
UTDOOR
/ T
ECH
/ P
RO
/ SECULIFEHIT
AM
/
MD
Fuse
FF (UR) 10 A/1000 V AC/DC,
10 x 38 mm,
switching capacity:
30 kA at 1000 V AC/DC,
protects the current input socket in the
100
A to 10 A ranges
Electrical Safety
per IEC 61010-1:2010/DIN EN 61010-1:2011/VDE 0411-1:2011
Safety class
II
Fouling factor
2
Test voltage
6.7 kV~
METRAHIT AM X-TRA / AM TECH / AM PRO / AM BASE
METRAHIT X-
TRA
/ O
UTDOOR
/ T
ECH
/ P
RO
/ B
ASE
Measuring category
III
IV
Operating voltage
1000 V
600 V
SECULIFEHIT
AM
/
MD
Measuring category
III
IV
Operating voltage
600 V
300 V
www.
.com
1.800.561.8187