Gossen MetraWatt METRAHIT 2+ Скачать руководство пользователя страница 16

16

GMC-I Messtechnik GmbH

13

Maintenance

Attention!

!

Disconnect the instrument from the measuring circuit 
before opening to replace batteries or fuses!

13.1 Battery

Make sure that no battery leakage has occurred before initial 
start-up, and after long periods of storage. Continue to 
inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely 
clean the electrolyte from the instrument with a damp cloth, 
and replace the batteries before using the instrument.
If the   symbol appears at the display , the batteries should 
be replaced as soon as possible. You can continue working 
with the instrument, but reduced measuring accuracy may 
result.
The instrument requires 2 x 1.5 V AA size batteries or two 
equivalent NiCd rechargeable batteries.

Replacing the Battery

Set the instrument face down onto a flat working surface, 
loosen the two screws at the back and lift off the housing 
base, starting at the bottom. The housing top and 
housing base are held together with the help of snap 
hooks at the top front.

Remove the batteries from the battery compartment.

Insert two new 1.5 V AA size batteries into the battery 
compartment, making sure that the plus and minus poles 
match up with the provided polarity symbols.

Important for reassembly: First set the housing base onto 
the housing top and align accurately (see photo below). 
Then press the two housing halves together, first at the 
bottom front (a), and then at the top front (b).

Secure the housing base with the two screws.

Please dispose of depleted batteries in accordance with 
environmental protection regulations!

13.2 Fuses

If one of the fuses blows, this condition is displayed at the 
LCD as soon as a measured quantity with a voltage of greater 
than 4 V is applied to the corresponding connector jacks. The 
following symbol appears at the digital display in this case: 

.

The 16 A fuse interrupts the 6 A and 10 A ranges, and the 
1.6 A fuse interrupts all other current measuring ranges. All 
other measuring ranges remain functional.
If a fuse should blow, eliminate the cause of overload before 
placing the instrument back into service!

Replacing the Fuse

Open the instrument as described under “Replacing the 
Battery”.

Remove the blown fuse with the help of an object such 
as a test probe, and replace it with a new fuse.

The following fuses are approved for use:

For current measuring ranges up to 600 mA:
Type FF 1.6 A / 1000 V AC (10 kA), 
6.3 mm x 32 mm

For 6 A and 10 A current measuring ranges:
FF 10 A/1000 V AC (30 kA), 
10 mm x 38 mm 

Attention!

!

Use specified fuses only! If fuses with other blowing 
characteristics, other current ratings or other 
breaking capacities are used, the operator is placed 
in danger, and protective diodes, resistors and other 
components may be damaged.

The use of repaired fuses or short-circuiting the fuse holder is 
prohibited.

13.3 Housing

No special maintenance is required for the housing. Keep 
outside surfaces clean. Use a slightly dampened cloth for 
cleaning. Avoid the use of cleansers, abrasives and solvents.

Device Return and Environmentally Compatible Disposal

The instrument is a category 9 product (monitoring and 
control instrument) in accordance with ElektroG (German 
Electrical and Electronic Device Law). This device is not 
subject to the RoHS directive. Furthermore, we make 
reference to the fact that the current status in this regard can 
be accessed on the Internet at www.gossenmetrawatt.com 
by entering the search term WEEE.
We identify our electrical and electronic devices in 
accordance with WEEE 2012/19/EU and ElektroG 
with the symbol shown to the right per DIN EN 
50419. These devices may not be disposed of with 
the trash. Please contact our service department regarding 
the return of old devices (see chapter 16).

(b)

(a)

Содержание METRAHIT 2+

Страница 1: ...Operating Instructions METRAHIT 2 Universal TRMS Multimeter 3 349 476 03 9 7 15 ...

Страница 2: ...elected 6 appears when acoustic signal is active 7 Selected type of current 8 Unit of measure 9 Unit of measure C F for temperature measurement 10 Triangle appears indicates overranging 11 Pointer for analog display 12 Δ REL relative measurement with reference to offset 13 Scale for analog display 14 Indicates that the negative analog display range has been exceeded 15 Warning regarding dangerous ...

Страница 3: ...intended purpose safety of the operator as well as that of the instrument is assured Their safety is however not guaranteed if the instrument is used improperly or handled carelessly In order to maintain flawless technical safety conditions and to assure safe use it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service and that ...

Страница 4: ...ernet at www gossenmetrawatt com by entering the search term WEEE Indicates EC conformity CAT III Maximum allowable voltage between the connector jacks and earth is 600 V category III DAkkS Calibration Upon Request See also chapter 14 Recalibration Repair Parts Replacement and Balancing When the instrument is opened voltage conducting parts may be exposed The instrument must be disconnected from t...

Страница 5: ...the measurement cables have been plugged in to the instrument 3 1 Automatic Measuring Range Selection The multimeter is equipped with automatic measuring range selection for all ranges except the 600 mV range Auto ranging is active as soon as the instrument is switched on The instrument automatically selects the measuring range which allows for highest possible resolution of the applied quantity T...

Страница 6: ...ous for observing measured value fluctuation and for balancing procedures The analog display has its own polarity indicator If the measured value exceeds the display range polarity at the analog display is switched after approximately 0 7 seconds Overranging is indicated by the triangle at the right 6039 digits 4 3 Display Illumination After the instrument has been switched on background illuminat...

Страница 7: ...o HOLD and MAN AUTO keys The instrument acknowledges zero balancing with an acoustic signal and 0 1 digit and Δ REL appears at the LCD depending on the measuring range The voltage displayed at the moment the key is pressed is automatically substracted from all subsequent measured values Zero balancing can be deleted by briefly pressing the MAN AUTO key which is acknowledged with an acoustic signal...

Страница 8: ...ent greater than 600 mA or the mA range for current less than 600 mA Activate the highest measuring range first when measuring current of an unknown magnitude Select the current type appropriate for the measured quantity by briefly pressing the FUNC key Each time the key is pressed DC and AC are alternately selected and switching is acknowledged with an acoustic signal The selected current type is...

Страница 9: ...applied to the corresponding connector jacks The following symbol appears at the digital display in this case If a fuse blows eliminate the cause of overload before placing the instrument back into service Refer to chapter 13 Maintenance regarding fuse replacement Zero Balancing Δ REL during Current Measurement You can select the reference point in the mA AC DC and A AC DC measuring range Plug the...

Страница 10: ...nation mV A The secondary terminals must thus be connected to and V 9 Resistance Measurement Make sure that the device under test is voltage free Interference voltages distort measurement results Set the rotary selector switch to Ω Connect the device under test as shown Zero Balancing Δ REL in the entire resistance measuring range Cable and contact resistances can be eliminated during resistance m...

Страница 11: ...tic Signal If the sub function continuity test with acoustic signal is activated a continuous acoustic signal is generated by the instrument within a display range of R 40 Acoustic Signal ON Briefly press the FUNC key Activation is acknowledged with an acoustic signal The symbol appears at the display as well Acoustic Signal OFF Briefly press the FUNC key once again Deactivation is acknowledged wi...

Страница 12: ...easured temperature in C at the digital display Briefly press the FUNC key to switch between C and F Note The cold junction temperature reference temperature is measured with a Pt100 temperature sensor inside the instrument It is displayed when the measuring input is short circuited Due to internal heating or a change from warm to cold environment or vice versa the reference temperature may differ...

Страница 13: ... 6 MΩ 50 pF 0 5 5 60 V 10 mV 5 MΩ 40 pF 4 4 MΩ 50 pF 0 5 5 600 V 100 mV 5 MΩ 40 pF 4 4 MΩ 50 pF 0 5 5 Voltage drop at approx range limit 5 A 60 mA 10 μA 100 mV 100 mV 1 0 5 10 D 1 5 5 10 D 1 0 A Cont A 600 mA 100 μA 700 mV 700 mV 1 0 5 1 5 5 10 D 6 A 1 mA 200 mV 200 mV 1 0 5 10 D 1 5 5 10 D 10 A 4 Cont 10 A 10 mA 300 mV 300 mV 1 0 5 1 5 5 10 D Open circuit voltage Meas current at range limit rdg d...

Страница 14: ... 0 1 50 200 C 1 K 2 200 400 C 1 2 Measured Quantity Frequency 30 Hz 45 Hz A 2 0 10 65 Hz 1 kHz 60 600 mA 6 A 1 5 10 10 A 2 10 30 Hz 45 Hz 600 mV 3 10 6 60 600 V 2 5 10 65 Hz 500 Hz 600 mV 3 5 20 65 Hz 800 Hz 6 60 V 2 5 10 Influen cing Quantity Sphere of Influence Measured Quantity Measuring Range Influence Error Battery Voltage 3 2 9 V 3 1 V 3 6 V V 2 Digits V 4 Digits A 4 Digits A 6 Digits 60 Ω 6...

Страница 15: ...oltage 600 V Fouling factor 2 Test voltage 5 2 kV per IEC 61010 1 2010 EN 61010 1 2010 Electromagnetic Compatibility EMC Interference emission EN 61326 1 2006 class B Interference immunity EN 61326 1 2006 EN 61326 2 1 2006 Fuses Fuse links for all ranges up to 600 mA FF UR 1 6 A 1000 V 6 3 mm x 32 mm switching capacity 10 kA at 1000 V with ohmic load protects all current measuring ranges up to 600...

Страница 16: ...orresponding connector jacks The following symbol appears at the digital display in this case The 16 A fuse interrupts the 6 A and 10 A ranges and the 1 6 A fuse interrupts all other current measuring ranges All other measuring ranges remain functional If a fuse should blow eliminate the cause of overload before placing the instrument back into service Replacing the Fuse Open the instrument as des...

Страница 17: ...nd answers By having your measuring instrument calibrated regularly you fulfill the requirements of a quality management system per DIN EN ISO 9001 15 Accessories 15 1 General The extensive accessories available for our measuring instruments are checked for compliance with currently valid safety regulations at regular intervals and are expanded as required for new applications Currently up to date...

Страница 18: ...etrology from test reports and factory calibration certificates right on up to DAkkS calibration certificates Our spectrum of offerings is rounded out with free test equipment management An on site DAkkS calibration station is part of our service department If errors are discovered during calibration our specialized personnel are capable of completing repairs using original replacement parts As a ...

Отзывы: