![Gossen MetraWatt METRAHIT 2+ Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/gossen-metrawatt/metrahit-2/metrahit-2_operating-instructions-manual_2247244013.webp)
GMC-I Messtechnik GmbH
13
12
Characteristic Values
1)
At 0
°
C ... + 40
°
C
2)
with zero balancing, + 35 digits without zero balancing
3)
without sensor
4)
12 A 5 minutes, 16 A 30 seconds
5)
1 ... 35 d from the zero point due to TRMS converter when probe tips are short-
circuited
6)
power limiting: frequency x voltage max. 3 x 10
6
V x Hz for U > 100 V
Key
rdg. = reading (measured value)
d = digit
Meas.
Function
Measuring Range
Reso-
lution
Input Impedance
Intrinsic Error at Max. Resolution
under Reference Conditions
Overload Capacity
1)
Meas.
Function
±
(... % rdg. + ... d)
±
(... % rdg. + ... d)
6000
5)
Value
Time
V
600 mV 100
μ
V
10 M
Ω
// < 40 pF
8.1 M
Ω
// 50 pF
0.5 + 5
1 + 5
600 V
DC
AC
eff
Sinus
Cont.
V
6
V
1 mV
5.2 M
Ω
// < 40 pF
4.6 M
Ω
// 50 pF
0.5 + 5
60
V 10 mV
5 M
Ω
// < 40 pF
4.4 M
Ω
// 50 pF
0.5 + 5
600
V 100 mV
5 M
Ω
// < 40 pF
4.4 M
Ω
// 50 pF
0.5 + 5
Voltage drop at approx. range limit
5)
A
60 mA
10
μ
A
100 mV
100 mV
1.0 + 5 (> 10 D)
1.5 + 5 (> 10 D)
1,0 A
Cont.
A
600 mA
100
μ
A
700 mV
700 mV
1.0 + 5
1.5 + 5 (> 10 D)
6
A
1 mA
200 mV
200 mV
1.0 + 5 (> 10 D)
1.5 + 5 (> 10 D)
10 A
4)
Cont.
10
A
10 mA
300 mV
300 mV
1.0 + 5
1.5 + 5 (> 10 D)
Open-circuit voltage
Meas. current at
range limit
±
(... % rdg. + ... d)
Ω
600
Ω
100 m
Ω
max. 1
V
max. 250
μ
A
1 + 5
2)
600 V
DC
AC
eff
Sinus
max. 10 s
Ω
6 k
Ω
1
Ω
max. 1
V
max. 100
μ
A
0.7 + 3
60 k
Ω
10
Ω
max. 1
V
max. 12
μ
A
0.7 + 3
600 k
Ω
100
Ω
max. 1
V
max. 1.2
μ
A
0.7 + 3
6 M
Ω
1 k
Ω
max. 1
V
max. 120
nA
0.7 + 3
40 M
Ω
10 k
Ω
max. 1
V
max. 50
nA
2.0 + 3
2
V
1 mV
max. 3
V
1 + 5
±
(... % rdg. + ... d)
600
Ω
0,1
Ω
< 1
V
1 + 5
600 V
DC
AC
max. 10 s
±
(... % rdg. + ... d)
°
C
TYP K
–50.0 ...
+400.0
°
C
0.1
°
C
1.0 + 5
3)
600 V
DC/AC
eff
Sinus
max. 10 s
°
C
±
(... % rdg. + ... d)
Hz
100 Hz
0,1 Hz
0.1 + 2
600 V
6)
Hz
1000 Hz
1 Hz
0.1 + 2