50
GMC-I Gossen-Metrawatt GmbH
Memory Occupancy Query:
OCCUP
Memory occupancy can be queried from the INFO menu.
Occupancy is read out to the main display in % from
00.00% to 99.99%.
SEt
V
inFo
↵
tiME
V
bAtt
V
tESt
rAM
V
OCCUP
↵
Exiting the Memory Mode via Menu Functions
➭
Select the main menu: StorE.
➭
Activate the
↵
key. Memory occupancy is displayed.
➭
Activate the
↵
key again, and StOP appears at the dis-
play.
➭
Activate the
↵
key once again, and the setup display
returns to SEt. MEM disappears. The memory mode is
deactivated.
➭
Press
ESC|FUNC
to return to the measuring function.
Exiting the Memory Mode via Shortcut
➭
Simultaneously activate the
ESC|FUNC
and
ON
keys.
rAM
CLEAR
– Delete Memory
Attention!
!
This function deletes all measurement values which
have been saved to memory.
The entire RAM can be cleared which contains, for exam-
ple, power disturbance data:
SEt
V
rAM
↵
no
V
YES
↵
22.1
General Parameters
Memory Duration:
durA
This parameter allows for a determination as to whether or
not measurement values should only be stored for a limited
amount of time. If memory duration is activated (on), a
period of time during which data are to remain in storage
can be entered in days and hours.
SEt
↵
V
durA
↵
OFF
V
on
↵
0–9 (dAYS)
VU
↵
00:00:00 (tiME)
↵
Toroidal Core Memory Mode:
CYCLE
rAM
If the toroidal core memory mode has been selected –
CYCLE
rAM
“on” – the oldest value is deleted and overwritten
with the new value when memory overflow occurs.
If
CYCLE
rAM
is set to “OFF”, the memory mode is deacti-
vated as soon as the memory is full.
The toroidal core memory mode cannot be activated if
rapid sampling has been selected (0.5 ms to 20 ms). The
selected setting is always perceived as “OFF”.
SEt
↵
V
durA
↵
V
OFF
↵
CYCLE
rAM
↵
OFF
V
on
↵