Insulation Resistance Measurement
at Telecommunications Equipment
with the METRA HIT 16T
Three jacks, a, b and j, have been provided for measurement at
telecommunications equipment with two conductors and shield.
Insulation testing between a and b, j and a or b and j can be
selected with the function selector switch.
➭
Connect the measuring cables to jacks a, b and j.
Caution!
!
Do not touch the conductive ends of the test probes
after insulation resistance measurement has been acti-
vated at the instrument.
If at all possible, only plug in the measurement cable
which is actually required for testing: loose test probes
or cable ends represent a safety hazard. A current with
a value of 1.5 mA (limited by the instrument) may flow
over your body, and although this is not life endanger-
ing, the electric shock is distinctly perceptible.
If you are taking a measurement at a capacitive DUT,
for example a cable, it may be charged with as much as
100 V. Touching the DUT after measurement may be
life endangering in such cases!
➭
Turn the selector switch to M
Ω
ISO
_a-b, then to M
Ω
ISO
_j-a and
finally to M
Ω
ISO
_b-j, in order to display possible interference volt-
age for all three possible conductor pair combinations.
Note!
Insulation resistance may only be measured at voltage-
free objects.
If an interference voltage of
>
50 V is present within the system,
insulation resistance measurement is disabled. The interference
voltage is displayed at the LCD. If a voltage of greater than 310 V
has been applied, this is indicated additionally with an acoustic
signal.
➭
To activate insulation resistance measurement: Briefly press the
multifunction key
FUNC
. Insulation resistance for the currently
selected conductor pair is displayed. Insulation resistance values
of less than 1 M
Ω
are indicated with an acoustic signal.
➭
Turn the selector switch to M
Ω
ISO
_a-b, then to M
Ω
ISO
_j-a and
finally to M
Ω
ISO
_b-j, in order to perform the desired test.
Automatic measuring range selection is active for insulation resis-
tance measurement. Their is no provision for the manual selection
of the measuring range.
600V
V
F
°
C
b
j
a
Measurement with METRA Hit 16T
Example:
Measurement of insulation
resistance at a 2 conductor
shielded cable
Insulation Resistance