Insulation Resistance Measurement
with the METRA HIT 16I/L
14.1 Preparation for Measurement
Note!
Insulation resistance may only be measured at voltage-
free objects.
When measuring high-ohmic insulation resistances do
not touch the measurement cables.
➭
Set the rotary switch to “V1M
Ω
”.
➭
Connect the measurement cables to the two jacks which have
been released by the automatic blocking system.
This selector switch position provides for the measurement of inter-
ference voltage.
➭
Turn the rotary switch to “M
Ω
ISO @1000V
” when the device under
test is voltage-free.
Note!
Position M
Ω
ISO @1000V
may only be selected for insula-
tion resistance measurement. Any inadvertently applied
interference voltage, however, is displayed in this posi-
tion.
If an interference voltage of
>
50 V is present within the
system, insulation resistance measurement is disabled.
The interference voltage is displayed at the LCD. If a
voltage of greater than 1000 V has been applied, this is
indicated additionally with an acoustic
signal.
High-Voltage!
Do not touch the conductive ends of the test probes
after insulation resistance measurement has been acti-
vated at the instrument.
A current with a value of 2.5 mA (limited by the instru-
ment) may flow over your body, and although this is not
life endangering, the electric shock is distinctly percepti-
ble.
If you are taking a measurement at a capacitive DUT, for
example a cable, it may be charged with as much as
1000 V, depending upon the selected nominal voltage.
Touching the DUT after measurement may be life
endangering in such cases!
Test Voltage Selection: 500 V or 1000 V
➭
If the
500V|1000V
key is briefly activated, the currently selected
test voltage is displayed.
➭
In order to select the other value, press and hold the
500V|1000V
key until the other value is displayed and acknowledged with an
acoustic signal.
600V
V
F
°
C
V
1M
M
1000V=ISO
R
ISO
Insulation Resistance Measurement