25
Maintenance
7
2. Disinfect surface of main unit and probe cord
- Clean the device gently with a clean, soft, lint-free cloth.
- Do not use solvents or any liquid when clean the device. The device is not waterproof.
- Do not use sharp or pointed tool to clean, as they may damage or scratch the device.
- Remove restorative material gently with plastic spatula. Avoid using sharp tools to clean the main unit.
1. Autoclaveable components
- File hook
- Lip hook
Recommended temperature and time: 135C (275F), 6 minutes minimum with a sterilization pouch.
Minimum drying time after sterilization: 10 minutes
- Do not sterilize in any way other than autoclave.
- Autoclave and dry at 135
℃
, but do not exceed that temperature.
- The file hook and lip hook must be thoroughly washed and cleaned before autoclaving.
- Any chemical or foreign debris left on instruments could cause them to malfunction or
could cause discoloration.
- It is highly recommended that instruments be autoclaved in a sterilization pouch
(wrapped) or similar device.
- Do not autoclave probe cord.
- Follow the manufacturer’s recommendations to disinfect files.
Autoclave file hook and lip hook after each patient.
Warning
Note
Sterilization
Warning
- In case change battery, please contact local dealer or Good Doctors.