* Depending on the angle of penetration of the x-ray beam, the apex may not appear correctly, and the
position of the apical foramen may appear to be located differently than it actually is.
4. Reading and radiography
5. After using the device
1) Turn the device off
2) Disconnect the probe cord and other cords or cables.
23
How to use
6
Sometimes the device meter reading and the x-ray image will not correspond.
This does not mean that device is not working properly or that the x-ray exposure is failure.
* Occasionally, the actual apical foramen does not correspond exactly.
The actual apical foramen may be located up towards the crown. In these case, the x-ray image will
seem to indicate that the file has not reached the apex.
Note
- Do not pull directly on the cords when connecting or disconnecting the probe and
file hook. Always grip the connectors to connect and disconnect cords.
- Do not wrap the probe cord around the body of the main unit.
X-ray Tube
X-ray Film
Apical foramen is located
up towards the crown.
GOOD
BAD