MAINTENANCE
72
Warning
To avoid electrical shock or arc blast, or personal injury or damage to the Meter,
use specified fuses ONLY in accordance with the following procedure.
To test the fuse: (See figure 13)
If the Meter does not respond when measuring current or transistor hFE, inspect to
see that the fuses aren’t broken.
To replace the Meter’s fuse: (See figure 12)
1.
Turn the Meter power off and remove all the connections from the terminals.
2.
Remove the screw from the tilt stand and the battery compartment and
separate the battery compartment and the tilt stand from the case bottom.
3.
Remove the two screws from the case bottom, and the separate the case top
from the case bottom.
4.
Remove the fuse by gently prying one end loose, then take out the fuse from its
bracket.
5.
ONLY install replacement fuses of an identical type and specification as shown
below and make sure the fuse is fixed firmly in the bracket.
µA/mA range:
F1, 600mA H 1000V,
φ
6.35x31.8mm (CE)
Содержание GDM-360
Страница 5: ...Table of Contents 3 Replacing the Fuses 71 USB and RS232C Serial Port 73...
Страница 24: ...MEASUREMENT OPERATION 22 C Resistance Measurement see figure 4 mAuA COM V Hz figure 4...
Страница 27: ...GDM 360 397 398 461 User Manual 25 D Testing for Continuity see figure 5 figure 5...
Страница 29: ...GDM 360 397 398 461 User Manual 27 E Testing Diodes see figure 6 figure 6...
Страница 32: ...MEASUREMENT OPERATION 30 F Capacitance Measurement see figure 7 mAuA V Hz figure 7...
Страница 37: ...GDM 360 397 398 461 User Manual 35 H Temperature Measurement GDM 397 only see figure 9 figure 9...
Страница 39: ...GDM 360 397 398 461 User Manual 37 I Transistor hFE Measurement GDM 398 only see figure 10 figure 10...
Страница 41: ...GDM 360 397 398 461 User Manual 39 J EF Function GDM 398 only see figure 11 figure 11...
Страница 73: ...GDM 360 397 398 461 User Manual 71 Replacing the Fuses figure 13...