Hardware and Software Components of Measuring
System
System Types
Page 6 (44)
at
os
2-
3_r
ev02_
so
-40
0-
800
-v7
-5
-0_en_r
ev
-b
2012
-01
-03
2.1.1
Technical Data
System
ATOS II Rev. 02
ATOS II Rev. 02 800
ATOS III Rev. 02
ATOS III Rev. 02 800
Measuring volumes for
camera position
800
560 x 420 x 420 mm to 2000 x 1500 x 1500 mm
(only for ATOS II Rev. 02 800 and ATOS III Rev. 02 800)
Measuring volumes for
camera position
400
170 x 130 x 130 mm to 1400 x 1050 x 1050 mm
Measuring volumes for
camera position
SO
38 x 29 x 15 mm to 320 x 240 x 240 mm
Measuring points per individu-
al scan
or camera resolution
approx. 5 000 000
(2448 x 2050 pixels)
approx. 8 100 000
(3296 x 2472 pixels)
Measuring point dis-
tance
800
0.23 to 0.79 mm
0.2 to 0.61 mm
400
0.07 to 0.62 mm
0.05 to 0.45 mm
SO
14.70 - 123.93 µm
11.58 - 95.46 µm
Transfer of image and control
signals
via Gigabit data link with max. 2 Ethernet interfaces
Projector light
Artificial optical radiation
blue, 400-500 nm
(Medium risk according to DIN EN 62471, issue March 2009,
when looking into the light source)
Ambient conditions
+5 °C to +40 °C
(non-condensing)
Voltage range (typical)
90 - 240 V, 50 - 60 Hz
Power consumption
Typically 130 W
max. 300 W
Power Factor Correction
≥ 95%
Max. cable length between
sensor and PC, laptop
30 m
For further information see http://www.gom.com
2.2
Hardware and Software Components of Measuring System
•
ATOS sensor with stereo cameras, fringe projector, controller and
power supply.
•
Stand for secure and steady hold of the ATOS sensor.
•
Please only use stands which are released by GOM
for this sensor type. Otherwise the stand with the sensor may fall
over!
•
High-performance PC system with Microsoft Windows operating
system and separate network interface card for the Gigabit data
link.
•
The factory settings of the separate network inter-
face card (interfaces Cam L, Cam R) are optimized for the Gigabit
data link and must not be changed! Never enable the firewall for
this network interface card
.
•
ATOS application software (Windows) as of version V7 see 2.1
Overview of the ATOS Triple Scan System Types.