Golden engineering XRS4 Скачать руководство пользователя страница 4

2

1.0  INTRODUCTION

 
      
The XRS4 is an industrial type open beam X-ray generator intended to radiograph inanimate objects.  The XRS4 is 
a pulsed X-ray device that produces X-ray pulses of very short duration (10 nanoseconds).  The energy produced 
by the XRS4 is up to 370kVp, which makes it possible to radiograph up to one (1.5) inch (3.81 cm) of steel.

XRS4 standard accessories are two keys, two battery packs, and one battery charger. Remote cable, carrying case, 
and film developing equipment are also common accessories. 

2.0   WARNINGS

The XRS4 is a pulsed X-ray generator that emits hazardous ionizing radiation when pulsing.  The XRS4 should only 
be operated by authorized personnel who are properly trained to safely operate the generator.  The XRS4 must 
be registered with proper authorities prior to use and should not be used to intentionally expose humans.      

Develop and closely follow a safe operating system for using the XRS4.  The safe operating system must ensure that 
no one is exposed to radiation above the permissible limits which are 2 mR (0.02 mSv) per hour for a member of 
the public.  

The safe operating system must ensure the XRS4 is used within federal and state guidelines. 

All operators and users of the XRS4 X-ray machine must wear a personal radiation monitoring device, such as a 
TLD (thermo luminescent dosimeter), film badge, and/or a pocket dosimeter consistent with the appropriate fed-
eral, territorial or provincial standards
 (note: an electronic dosimeter will not detect the XRS4 radiation pulses).  If 
an operator or bystander is exposed to an unacceptable level of radiation contact your Radiation Safety Officer 
and/or appropriate health care provider. 

Due to the short pulse width of the XRS4, survey meters of the Geiger-Mueller and scintillator type do not accu-
rately detect the radiation emitted from the X-ray source.  

Survey meters should be of the ionization type and should be used in the integration mode

.  Survey meters must not 

be used in the rate mode because the XRS4 does not produce constant radiation.  The XRS4 produces very high 
rates of radiation for very short periods of time resulting in either unrealistically high readings or no readings for 
a survey meter in rate mode. 

The XRS4 has no explosion proof rating and should not be used in an explosive atmosphere. The Spark Gap 
is vented to the air and could be a source of ignition.  

THE STANDARD XRS4 WITHOUT A MODIFIED SPARK 

GAP WILLL LIKELY FAIL PREMATURELY WHEN USED ABOVE 5000 ft. (1524 m) ELEVATION.   

  

The XRS4 produces high levels of radiation and must be operated by qualified personnel who have read 

the Warning and Operations section of the manual before operating the device.

DUTY CYCLE WARNING.  

The XRS4 is a light duty machine that is not made to pulse continuously.  

The maximum 

duty cycle for the XRS4 is 200 pulses every four minutes (3000 pulses per hour).  Additional 30 second rest every 
50 pulses is suggested for applications requiring more than one hour continual use or in temperatures exceeding 
90 degrees F.  Exceeding the duty cycle will shorten the life of the tube and head.

Содержание XRS4

Страница 1: ......

Страница 2: ......

Страница 3: ...AVIGATING THE MENU 10 SET PULSE COUNT 11 TIME DELAY OPTION 12 MULTIPLE PULSE TRAIN OPTION 13 DUTY CYCLE 14 PULSE SETTINGS 14 6 0 MAINTENANCE 15 DOSE MEASUREMENT 15 TUBE REPLACEMENT 15 7 0 TROUBLE SHOOTING 15 8 0 INSTRUCTIONS FOR REPAIR 16 FUSE REPLACEMENT 16 REMOVING BOARDS 17 BOARD INSTALLATION 17 STORAGE TRANSPORT DISPOSAL 17 9 0 WARRANTY 18 MANUFACTURER EUROPEAN REPRESENTATIVE 18 RETURNING REPA...

Страница 4: ... note an electronic dosimeter will not detect the XRS4 radiation pulses If an operator or bystander is exposed to an unacceptable level of radiation contact your Radiation Safety Officer and or appropriate health care provider Due to the short pulse width of the XRS4 survey meters of the Geiger Mueller and scintillator type do not accu rately detect the radiation emitted from the X ray source Surv...

Страница 5: ... and a threaded 20 insert that can be attached to any standard camera tripod Identification label indicates the model manufacturer and serial number is located on the bottom of the XRS4 base BATTERY PACK The standard battery pack is a DeWalt 18V nickel cadmium battery Optional battery chem istry or voltage may be available BATTERY CHARGER The standard battery charger is the DeWalt DW9116 110V char...

Страница 6: ...ltiple pulse trains Time Delay DEL and Ready to Fire See instructions for menu options DELAY SWITCH Initiates time delay mode count down when left and right arrows pressed simultaneously POWER ON LIGHT Illuminates when battery voltage is applied to control module ENTER EMERGENCY STOP SWITCH ENTER used when menu options are selected EMERGENCY STOP stops the unit during delay count down mode or betw...

Страница 7: ...puts are activated when grounded 3 4 5 2 1 CABLE CONNECTOR Lemo K series five pin connector located on the back of the control module beneath the battery receives the remote cable or imaging system cable PIN DESCRIPTION 1 5 VOLTS 100 ma MAXIMUM 2 REMOTE SWITCH 3 REMOTE SWITCH NO DELAY 4 X RAY ON SIGNAL 5 COMMON 0 VOLTS Remote cable connector ...

Страница 8: ... pulsed 7 As the High Voltage Switch is closed a high voltage transient of approximately 370 000 volts and 10 nanoseconds in duration is applied across the X ray tube generating X rays The closing of the High Voltage Switch produces an audible pulsing sound The XRS4 cannot produce x rays with out the pulsing sound so it serves as an additional warning the XRS4 is functioning This unit generates X ...

Страница 9: ...ey and turning clockwise 1 4turn Turn on XRS 4using key and turning clockwise 1 4turn Set Pulse Count pg 11 Set pulse count to 200 pg 11 Set Pulse Count pg 11 Set Delay Time pg 12 Retreat to imaging system controller computer Retreat length of cable behind the XRS 4 Press left and right buttons simultaneously to initiate time delay Follow imaging system instructions to set pulses and initiate puls...

Страница 10: ...urface a tripod or a custom fixture suitable for holding the 22 pound 9 Kg XRS 4 OPERATING PRECAUTIONS The operator should always stand at least 20 feet 6m behind the X ray unit and clear all personnel at least 20 feet 6m behind the unit or at least 100 feet 30m from the front of the unit before pulsing The exclusion zone below should be a controlled area free of all personnel while X ray pulses F...

Страница 11: ...film is underexposed If the radiograph is too light the film is overexposed Underexposure can be cor rected by increasing the number of pulses and or decreasing the distance between the imaging medium and the XRS4 Overexposure can be corrected by reducing the number of pulses and or increasing the distance between the imaging medium and XRS4 MATERIAL PULSE SETTING 24 BETWEEN X RAY IMAGER CARDBOARD...

Страница 12: ...ATING THE MENU Home Press Right Arrow to advance to next menu Press Left Arrow to go back to home Press enter to view settings menu options Down Arrow to scroll to next menu option Press enter to select option ...

Страница 13: ...each pulse train Press Right Arrow button from home screen Press Enter button from Settings screen Select Set Pulse Count and press Enter Button Use LEFT RIGHT arrow to select Digit Use UP DOWN arrows to change number Pulse setting from 001 to 999 Press Right Arrow to Save and Press enter Home screen appears after pulse count is saved ...

Страница 14: ...e trains and does not affect delay time before the first pulse when using a remote cable or when X ray is fired from a digital imaging system controller Press Right Arrow button from home screen Press enter button from Settings screen Select Set Delay Time press Enter button Use Left Right arrows to select digit Use UP DOWN arrows to change number Delay option from 5 to 240 seconds Press Right Arr...

Страница 15: ... a single pulse train This safety fea ture ensures operator s knowledge that multiple pulse trains are set Red X ray pulse light blinks between multiple pulse trains Multiple pulse train option can be used slow the pulse rate to one pulse per second by entering a pulse count of one enter desired number of pulse trains pulses and select one second between pulse trains Select Multiple Trains press E...

Страница 16: ...pair Golden Engineering can see how many times the duty cycle had been exceeded If the repairs required to the machine are caused by excessive use the warranty may be voided PULSE COUNTERS XRS4 contains two pulse counter options The first pulse counter can be reset to zero by selecting Pulse Counter then pressing the UP ARROW to reset This feature allows the operator to track the number of pulses ...

Страница 17: ...he tube is broken or the glass cracks the tube output will cease immediately 7 0 TROUBLESHOOTING SYMPTOM TEST ACTION No power on light Check battery voltage Check battery connection Replace or charge battery Make sure battery is securely attached and battery clips are not bent or broken Power on lights but X ray does not pulse Check the battery voltage Check the 2amp fuse Charge or replace the bat...

Страница 18: ...the opening in the oscillator board 2 The 20 amp fuse is the white one inch long fuse on the left side of the oscillator board The 15 amp fuse is a small green fuse in the center of the oscillator board The 15 amp fuse is soldered in place and can only be replaced using a soldering iron 3 REMOVING THE BOARDS Refer to the diagram on page 17 4 Using T10 Torx Remove the 2 screws on both sides of the ...

Страница 19: ... oscillator board in place Two are in the middle of the board and one is at the bottom Remove these three screws then remove the Oscillator Board BOARD INSTALLATION 1 Install the Oscillator Board 2 Put the three cap screws through the oscillator board and then put the offsets on the back of the screws 3 Attached the two blue feedback wires one red feedback wire and green signal line 4 Attach the C...

Страница 20: ...t customer s expense THIS EXPRESS LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES AND GUARANTEES EITHER EXPRESS OR IMPLIED OR CREATED BY OPERATION OF LAW XRS4 Manufacturer European Representative Golden Engineering Inc Certification Experts Europe PO Box 185 Nieuwstad 100 Centerville IN 47330 USA 1381 CE Weesp Phone 1 765 855 3493 The Netherlands Fax 1 765 855 3492 Web www goldenengineering co...

Страница 21: ...el Cadmium sub C cells Battery recharge time One Hour Current draw 13 amps 18 volts Current flow 0 25 mA Temperature range 10 to 120 degrees F 23 to 50 degrees C Airborne Noise Emissions 80 dB at 10 cm Maximum duty cycle 200 pulses every 4 minutes 3000 pulses per hour High temperature High use duty cycle Above 90 Degrees OR continual use rest 30 sec onds every 50 pulses and 4 minutes every 200 pul...

Страница 22: ......

Страница 23: ......

Страница 24: ......

Отзывы: