4.8 Configuring the Sensor Modules
4.8.1 The Parameter Window
RS485 / CAN Termination
Switches the termination of the Modbus (RS485) and the CAN bus on/off.
Applies only to older BlueConnect Modules, the newer ones are terminated with slide switches on the
board, see
3.2 Cable Connections, Switch Positions and LEDs
there also
Note on termination of older BlueCon-
nect Modules
. Newer modules with slide switches ignore the setting.
O2
Only visible with O2 Sensor Modules.
Selection
mg/l
or
% Saturation
This selection determines the type of calibration (see
4.8.2 The Calibration Window O2
) and how
the measurement value is stored and displayed
Coefficient O2
Calibration coefficients, the displayed values are from
the Calibration function, see
4.4 The Calibration Window
.
Coefficient Temperature
Only visible with Sensor Modules.
Factory calibration coefficients of an assigned temperature sensor
∗
.
If necessary, you can determine the offset here via the Coefficient A0.
Writes the input settings into the module memory.
Settings that have not yet been saved are marked in red.
Note:
Decimal separator is the comma; if a dot is entered, an error message appears.
∗
In this case, the internal temperature sensor of the O2 Sensor.
COM 1 selected
Serial Communication
Modbus
BlueConnect O2
Info
Parameter
Calibration
Measuring
Data Processing
Data
RS485 / CAN Termination on o ff
O2 mg/l %
write changes
Coefficients O2
Coefficient A0
A1
Pressure A2
Salinity A3
A4
A5
Coefficients Temperature
A0
A1
A2
A3
A4
A5
-4,975610E-01
1,00
-9,711752E-02
0,00
0,00
0,00
-1.01
5.594206E-02
-3.445109E-05
1.625741E-08
-3.872879E-12
3.711060E-16
File Language
Modbus Tool V1.07
write changes