background image

English

13

DEAR CUSTOMER

Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY

product comparable to the best products available in the World. This manual is a general presentation of
your product and it will give you important details to help you to solve possible problems.

Please, read it and keep this with you.

PRODUCT DESCRIPTION

The Optilight LD Max belongs to the newest generation of 

LED 

photo-activation devices. This

abbreviation stands for 

Light Emitting Diode

, a totally different type of light emission, if compared to

conventional halogen equipment.

Unlike traditional devices, which generate wide-spectrum light and heat, this technology uses a cold

light of the precise wave length needed to activate various dental products.

LED technology, which was recently introduced in Dentistry, brought about several useful features to

those light-curing devices used in composite resin restoration. Besides being more durable, LED technology
turned devices more compact, ergonomic and easier to install and transport. The emission of cold light
within a precise wave length range ensures the safe cure of camphorquinone-activated composites,
preventing dental heating, pulp damage or discomfort for both patient and dentist. Although being relatively
new, this technology is nowadays in its second generation. LED safety and efficiency, now allied to high-
energy emission, are available to all clinic procedures which require light-curing power, including bleaching
treatments.

The light of 440nm-460nm wave length, allied to the high energy emitted by Optilight LD Max, makes

possible the multi-functionality of this device:

-

Direct restoration procedures

: composite resins, ionomers and adhesives.

-

Indirect restorations

: adhesive cementation of laminates, inlays, esthetic pins and metal-free

crowns.

-

Dental Bleaching

: activation of bleaching gel and polymerization of gingival barriers. Compatible

with 35% hydrogen peroxide-based bleaching gels.

- Attachment of braces and orthodontic accessories.
-

Activation of light-cure materials

, such as sealants, surgical cements and covering bases.

Designed and built with cutting-edge technology, it meets the highest standards specified by world’s

dental authorities.

It features an automatic bi-voltage power supply, which allows operation in the range of 90 to 240 V

AC, 50/60 Hz.

Operation control display in handpiece, sound alarm with beep every 10 seconds and 4 beeps at the

end of the cycle.

Advantages offered by Optilight LD Max:
- More spectrally-selective light than conventional lamps.*
- Cold light, it doesn’t heat up the resin nor the tooth**
• Light compact equipment that provides handling comfort.
• Low power consumption.
• Longer useful life of the light emitting diode (equivalent to 36.000.000 cycles of 10 seconds).
• It does not use optical filter.
• It does not require forced ventilation, thus avoiding noise emission.

* We noted that the light emitted by the Optilight LD Max is completely contained within the absorption

interval of the photo starter, therefore it’s 100% used, whereas the conventional equipment running on
halogen lamps has non-used wave-length regions.

** The Optilight LD Max doesn’t generate heat since it uses light emitting diodes.
The light conductor is removable, made out of high resistance polymer and of easy maintenance.
The handpiece’s low weight and its anatomic design ensure the dentist more comfortable and practical

handling.

Содержание Optilight LD Max

Страница 1: ......

Страница 2: ...possible the multi functionality of this device Direct restoration procedures composite resins ionomers and adhesives Indirect restorations adhesive cementation of laminates inlays esthetic pins and metal free crowns Dental Bleaching activation of bleaching gel and polymerization of gingival barriers Compatible with 35 hydrogen peroxide based bleaching gels Attachment of braces and orthodontic acc...

Страница 3: ...res Products manufactured in compliance with RDC 59 ANVISA Sanitary Surveillance NationalAgency which guarantees compliance with sanitary regulations concerning BPF Good Manufacturing Practices requested by Mercosul Sanitary authorities Note Use handpieces and accessories that comply with requirements of directive 93 42 EEC with CE mark IMPORTANT This equipment is exclusively for dental use It mus...

Страница 4: ...ower Supply 02 Cable PP 2 20m 03 Optilight LD Max body 04 Pen 05 Ocular protector 06 Light conductor 07 Light conductor Whitening Tip 08 Display 09 On Off timer button 10 Time selection Whitening Tip 08 01 02 06 05 04 03 07 09 10 ...

Страница 5: ...esis orthodontic device implants and many other applications that guarantees its bio compatibility with the mouth mucus Applied norms This product was tested and approved in compliance with the following standards EN 60601 1 1990 Amendment 1 EN 60601 1 1992 Amendment 2 EN 60601 1 1995 Amendment13 EN 60601 1 1995 EN 60601 1 2 2001 NBR IEC 601 1 1994 Emenda 1 NBR IEC 601 1 1994 NBR IEC 601 1 2 1997 ...

Страница 6: ...peratures from 12ºC to 50ºC With maximum piling up of 13 units for storage Content of accessible and non accessible demarcations Special providence or particular conditions for installation Check the electric net it must be compatible with the specified one in the equipment Verify the general key of the equipment in the position 0 OBS These information also make part of the Manual of Installation ...

Страница 7: ...or stored with care should not suffer drop and neither receive impact The packing must be stored and trans ported away from direct sun light exposure Temperature limit for the packing to be stored or transported Packing symbols TECHNICAL SPECIFICATIONS Product symbols Warning Consult the manual Landing in many parts of the equipment indicates the condition of being landed B type equipment Time sel...

Страница 8: ...English 19 Dimensions mm TECHNICAL SPECIFICATIONS ...

Страница 9: ...ght conductor tip 06 at least 2mm away from the restoration Keep the light conductor 06 always protected by an expendable PVC film which must be changed for every patient This procedure protects the light conductor from scratches and other residues Use the polymerization time recommended by the compound resin manufacturer and always perform restorations in incremental layers with a maximum thickne...

Страница 10: ...LE AND OR REMOVABLE PARTS REPLACEMENT If stains fissures or scratchs where detected in the eyesight protection or in the light conductor replace damaged components Cleaning conductor The light conductor cleaning and the optical protector must be done using only neutral soap and cotton To the exterior of the pen use neutral soap or alcohol 70 vol Never use any other chemical based product than prev...

Страница 11: ...pment Doubts and information GNATUS Call center 55 16 2102 5000 TROUBLESHOOTING FINAL CONSIDERATIONS ATTENTION Any other problem that could happen with its equipment should only be repaired by an authorized technician Gnatus The bad use negligence or maintenance not made by technician authorized by the maker will imply in the loss of the warranty Imprevisto Problem Probable Cause Solution Equipmen...

Отзывы: