.Fault resolution

 

FAULTS 

REASONS 

RESOLUZTIONS 

 

Data 
inaccurate   

 

1. On-site calibration method is not 
correct. 
2. Factory calibration is not suitable for 
liquid under measurement. 
3. Instrument operation is less than the 
minimum flow rate. 
4. The flow meter is blocked by part of 
dry liquid. 
5. Sealant material wrapped around 
the impeller.   
6. Packed with accessories too closely. 
7. Installed too close to the motor or 
electronic noise environment. 

1. Re-calibration on-site or scene 
selection for Factory calibration.   
2. Calibration on-site by Check   
Chapter 
3. Remove the impeller and carefully 
cleaned. To ensure that the impeller 
rotating freely.   
4. Removed instrument ( LCD 
screen ) to ensure that the impeller 
rotating freely.   
5. Installed correctly.   
6. Installed correctly. 

 

Readings are 
degenerating 
or blanking 

1, Power of batteries is weak, or even 
on depleted.   
2, The computer failure. 

1. Removed the computer and, if 
necessary, check and reinstall or 
replace the batteries.   
2. Factory contact. 

Normal flow 
rate but the 
instrument 
does not 
count, when 
pressing the 
DISPLAY 
button, the 
instrument 
boots. 

1, On-site calibration method is not 
correct.   
2, Impeller jammed or damaged.   
3, Sealant material wrapped around 
the impeller.   
4, The computer failure. 

 

Re-check on-site, or check by 
Factory.   
Removed instrument to ensure that 
the impeller rotating freely.   
Removed instrumentation to ensure 
that the impeller rotating freely.   
Contact factory   

 

The flow rate 
meter reduced 
and can not 
counter, when 
press 
“DISPLAY “, 
instrument 
boots. 

1

The flow meter is blocked by dry 

liquid. 
2

Lower than minimum flow rate 

Removed instrument ( LCD screen ) 
to ensure that the impeller rotating 
freely. 

 

Improve flow rate. 

Instrument is 
not allowed 
the access to 
the site check.

 

 

1, The computer circuit board failure.   
2, Buttons failure. 

 

Replace the computer, contact 
factory 
Replace the computer, contact 
factory 

 

Tel:86-21-57632436                                                   Fax:86-21-57634942

Shanghai Cixi instrument Co., ltd

www. gnflowmeter.com

Содержание CX-WLTFM

Страница 1: ...Usermanual CX WLTFM TURBINEFLOWMETER CX WLTFM TURBINEFLOWMETER Tel 86 21 57632436 Fax 86 21 57634942 Shanghai Cixi instrument Co ltd www gnflowmeter com ...

Страница 2: ...flow coefficient k and user manual calibration of the flow coefficient k calibration 2 Functions as below Two cumulate which can be reset TTL1 cumulate Clear TTL2 the total accumulation can not clear zero it can be shifted by the buttons Unit selection The selection buttons to operate the flow units liters and gallons U S Sensor pulse frequency range 5Hz 5000Hz Low power consumption Two pieces of ...

Страница 3: ...ss CALIBRATE can shift the unit LTR and GAL and the corresponding characters will be showed on the LCD display To press DISPLAY and CALIBRATE at the same time about 2 seconds will enter the user manual flow coefficient k calibration status B LCD Description TTL1 current flow can be cumulative in which can be cleared zero TTL2 current flow displayed is total cumulative in which can not be cleared z...

Страница 4: ...ut about 30 seconds control panel will record the flow value calculate the flow coefficient k values and deposited CPU of the permanent preservation of non volatile memory even removed the battery discharge coefficient k will not lose value If the flow of calibration error over more than the limited 10 the panel will consider the current calibration value is invalid and will not record the current...

Страница 5: ...0 L k l k Note l flow rate of flow meter k0 current flow not to amend before k value L0 user graduated cylinder measures the value Users to manually modify the k value for example Clear 0 TTL1 value or remove the batteries to clear TTL1 TTL2 0 fitted with batteries button is shifted to TTL1 status start the measurement of liquid flow the flowed liquid enters into the standard cylinder Record the d...

Страница 6: ...tenance Battery electrode should be inspected once a year No need to remove the instrument from the pipeline system the batteries can be replaced V Batteries replacement Remove the four angle screws from the instrument s control panel take the computerized electronic instruments LCD screen from the turbine flow Take out the batteries Check battery electrode if rust and clear it Install new batteri...

Страница 7: ...ilure 1 Removed the computer and if necessary check and reinstall or replace the batteries 2 Factory contact Normal flow rate but the instrument does not count when pressing the DISPLAY button the instrument boots 1 On site calibration method is not correct 2 Impeller jammed or damaged 3 Sealant material wrapped around the impeller 4 The computer failure Re check on site or check by Factory Remove...

Отзывы: