![GN Resound Pulse CRT LT PSS60-R Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/gn-resound/pulse-crt-lt-pss60-r/pulse-crt-lt-pss60-r_technical-manual_2236521010.webp)
Doc. No. 0127420 rev. B
Page 10 of 15
Hybrid connections
Contact module
Battery springs
BA
T(+)
H8
H9
/REC(+)
REC(-)
MIC+
MIC-
MIC IN
Contact
module
BA
T(-)
Contact
module
Coyote 3.0 - No marking
Coyote 3.1 - Black dot marking
The layout of C3.0 and C3.1 is the same. For visual purpose the C3.1 is
marked with black dots. Only PS60-R is mounted with hybrid C3.1. In a
repair situation a hybrid C3.0 can be replaced to C3.1. The test system
will automatically detect the hybrid version during test.
NOTE:
Please also see Technical News Bulletin No. 2007-08-001 (available
from GTOweb) which is describing the implementation of C3.1.
Furthermore the improved test fixture is described