WARNINGS 

 

REFER TO WARNINGS BELOW BEFORE OPERATING ELECTROMAGNET SYSTEM 

 

ELECTROMAGNET 

1.  Safety 

In operation the magnet fringing field in the vicinity of the pole gap is in excess of 0.5mT (5G).  This can cause 
malfunctioning of sensitive electronic and magnetic components.  We recommend that warning signs are posted 
indicating that a magnetic field may be present. 

 

2.  Ferromagnetic Objects

 

During operation the magnet exerts magnetic attraction towards ferromagnetic objects in the near vicinity of its 
pole gap.  Keep ferromagnetic items clear! 

 

3.  Arcing

 

This magnet stores energy in its field during operation. Do not disconnect any current lead while under load or 
the magnetic field energy will be discharged across the interruption causing arcing and possible damage to 
electronic circuits. 
 
4.  

Coil Hot Resistance

 

Do not exceed the maximum coil hot resistance given in the specifications or coil overheating and possible 
damage may occur. 

 

5.  Watches, Credit Cards, and Magnetic Disks 

 

Do not move magnetically sensitive items into the close vicinity of the magnet pole.  Even some anti-magnetic 
watches can be damaged when placed in close proximity to the pole gaps during operation. Credit cards, and 
magnetic disks are affected by magnetic fields as low as 0.5mT (5G). Depending on the previous operating 
field and the pole gap, the remanent field in the gap can be in excess of 0.5mT (5G) with the magnet power 
supply off or disconnected. 

 

    --------------------------------------------------------------------------------------------------------- 

 

MOTORIZED ROTATIONAL DRIVE 

 

1.  Hand Operation 

Do not rotate the Motorized Rotating Drive platen by hand. For hand operation the Motor Drive assembly 
must be removed. See maintance section for instruction on how to remove the Motor Drive assembly. 
 

2

.  

Obstructions  

Keep foreign objects  and tools  clear of the Motorized Rotation Drive. Ensure the Electromagnet and 
Motorized Rotation Drive do not contact obstructions during rotation. 
 
 
 
 

2-1 

Содержание 5201

Страница 1: ...OT BE REPRODUCED OR DISCLOSED TO OTHERS OR USED IN ANY WAY EXCECPT FOR THE INSTALLATION OPERATION OR MAINTENANCE OF GMW ASSOCIATES PRODUCTS File No M5201a Projected Field Electromagnet doc Revision Date April 10 2003 _____________________________________________________________________________________ GMW 955 Industrial Road San Carlos CA 94070 Tel 650 802 8292 Fax 650 802 8298 Email sales gmw com...

Страница 2: ...ersus Current TEST DATA Section 7 Field Uniformity DRAWINGS Section 8 Drawing 11901960 Electromagnet Motorized Rotation Drive General Assembly Drawing 11901860 Electromagnet General Assembly Drawing 11901950 Electromagnet Electrical Assembly Drawing 13900420 Electromagnet Electrical Wiring Drawing 13900430 Electrical Wiring Kepco Programming Plug Drawing 11901910 Electromagnet Terminal Block Assem...

Страница 3: ...e hot 1 02 Ohm max power 20A 20V 400W Self Inductance Cooling 0 5 Litre min 1 0 bar 14 psid Thermal Interlock Open circuit above 75 C 167 F Dimensions Drawing 11901860 70 0 mm W x 60 0 mm D x 120 mm H 2 8 inch W x 2 4 inch D x 4 7 inch H Weight 4 kg 9 lb CAUTION The value of maximum coil resistance given should not be exceeded At this resistance the coils are at maximum safe temperature for contin...

Страница 4: ...x 200 deg max Homing Switch 0 0deg Drive System Gear type Worm Worm Gear Mechanical reduction 100 1 Stepper Motor Motor type VS13B SFRIO Motor torque 0 5Nm 73 oz in Motor frame size 17 Stepper Motor Controller Model Compumotor Zeta 6104 AC Power 115V AC only Rotation Axis Scaling approx 6944 steps deg Dimensions Drawing 11901960 200 mm W x 200 mm D x 193 mm H 7 9 inch W x 7 9 inch D x 7 6 inch H W...

Страница 5: ...ed the maximum coil hot resistance given in the specifications or coil overheating and possible damage may occur 5 Watches Credit Cards and Magnetic Disks Do not move magnetically sensitive items into the close vicinity of the magnet pole Even some anti magnetic watches can be damaged when placed in close proximity to the pole gaps during operation Credit cards and magnetic disks are affected by m...

Страница 6: ... plug with the two securing thumbscrews 3 Connect the three sleeved wires to the output connector block on the rear of the Kepco BOP power supply as detailed below Black Wire with RED sleeve to Output Black Wire with BLUE sleeve to Common Green wire to Ground 4 Connect GPIB ribbon cable to edge connector on GPIB board 5 Connect GPIB adaptor into the back of Kepco BOP power supply 6 Secure GPIB ada...

Страница 7: ...r will trip to the off position Clean cool 16 C 20 C water at 0 5 l min at 1 0 bar 14 psid should be used to cool the 5201 magnet The cooling copper tubes are electrically isolated from the coils to avoid electrochemical corrosion A 50 micron filter should be placed before the input to the magnet to trap particulates and avoid blockage of the cooling circuits Water Cooling Connections The magnet s...

Страница 8: ...stall the driver follow the directions given below 1 Insert CD rom into computer CD ROM drive 2 Copy the complete 5201 Magnet Control directory to the host computers hard drive 3 Locate the Default dat file in the Data directory 4 Right mouse click on the file properties 5 Uncheck the Read Only attribute Before running the 5201 Magnet Control software set the Com Port control to match the Com Port...

Страница 9: ...om the magnet power supply Loaded Excitation File Displays the current loaded excitation file MAGNET ROTATION CONTROL PANEL Home CW Control Makes the motorized rotating Drive rotate in a clockwise direction until it finds the homing position at 0 0 deg and resets the current position counter to 0 0 deg Home CCW Control Makes the motorized rotating Drive rotate in a counterclockwise direction until...

Страница 10: ...ets the acceleration deceleration and velocity of the stepper motor when homing to the 0 0 deg position Soft Limits Range Sets the maximum minimum and increment on the soft limits scale Soft Limit feature can be turned On or Off by the Soft Limits control switch Velocity Chart Indicator Shows the current acceleration deceleration and velocity of the stepper motor Reset Controller on Start Up Sends...

Страница 11: ...og file Auto log indicator will flash yellow CREATE DATA FILE PANEL Visible only when the new data file control is pressed Data File Header Control Provides inputs for test engineer magnet model magnet serial no andtest notes The computer date is automatically inserted into the data file Data File Name Control Provides input for new file name Create New Data File Control Creates a new data file an...

Страница 12: ...rol or indicator on the front panel A description of the control or indicator function will appear in the help window ELECTROMAGNET OPERATION Current Control 1 Select current control mode default 2 Select Setpoint on the key focus control 3 Enter desired current from the keyboard and press the enter key The setpoint control pointer should move to the desired current and the correct current will be...

Страница 13: ...ion Soft Limits Operation 1 Set the soft limits control to enable 2 Set the CCW limit by clicking on the blue pointer with the mouse and dragging to the desired location on the status indicator 3 Set the CW limit by clicking on the blue pointer with the mouse and dragging to the desired location on the status indicator 4 Click on either the CCW or CW Start Motion button The motorized rotating driv...

Страница 14: ...n is overheating and must be rectified before further use of the magnet Water Hoses should be checked regularly for water leaks Any leaks should be rectified before further use of the magnet Motorized Rotating Drive Rotating Platencontains nylon ball bearings and does not require periodic lubrication Worm Gear should be lubricated annually by smearing the worm gear with a coating of a graphite bas...

Страница 15: ...Section 6 EXCITATION CURVES ...

Страница 16: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 2mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 17: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 3mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 18: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 4mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 19: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 5mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 20: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 6mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 21: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 7mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 22: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 8mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 23: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 9mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 24: ...Projected Field Magnet SN 001 Bx vs Current X Y 0mm Z 10mm 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 20 15 10 5 0 5 10 15 20 Current A Field T Upward Ramp Downward Ramp ...

Страница 25: ...Section 7 TEST DATA ...

Страница 26: ...Projected Field Magnet SN 001 Bx vs Z X Y 0mm Current 20A 0 00 0 05 0 10 0 15 0 20 0 25 0 30 0 35 0 40 0 45 0 50 0 1 2 3 4 5 6 7 8 9 10 11 12 Z mm Bx T ...

Страница 27: ...ted Field Magnet SN 001 Bx vs X Y 0mm Current 20A 0 00 0 05 0 10 0 15 0 20 0 25 0 30 0 35 0 40 0 45 0 50 5 4 3 2 1 0 1 2 3 4 5 X mm Field T Z 2mm Z 3mm Z 4mm Z 5mm Z 6mm Z 7mm Z 8mm Z 9mm Z 10mm Z 11mm Z 12mm ...

Страница 28: ...ed Field Magnet SN 001 Bx vs Y X 0mm Current 20 A 0 00 0 05 0 10 0 15 0 20 0 25 0 30 0 35 0 40 0 45 0 50 5 4 3 2 1 0 1 2 3 4 5 Y mm Field T Z 2mm Z 3mm Z 4mm Z 5mm Z 6mm Z 7mm Z 8mm Z 9mm Z 10mm Z 11mm Z 12mm ...

Страница 29: ...ted Field Magnet SN 001 Bz vs X Y 0mm Current 20A 0 50 0 40 0 30 0 20 0 10 0 00 0 10 0 20 0 30 0 40 0 50 5 4 3 2 1 0 1 2 3 4 5 X mm Field T Z 2mm Z 3mm Z 4mm Z 5mm Z 6mm Z 7mm Z 8mm Z 9mm Z 10mm Z 11mm Z 12mm ...

Страница 30: ...Section 8 DRAWINGS ...

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ...Section 9 SOFTWARE REVISION LIST The Software revision number has the following format V1 1a Letter denotes bug fix Major version number Minor update number V1 0 First release April 10 2003 ...

Отзывы: