
4-5
INSTALL INSTRUMENT IN STATION & PERFORM BUMP / CALIBRATION
Close the front cover of the Auto Bump / Cal Station, as
illustrated in Fig. 4-6, then, using both thumbs, press the lower
edge corners of the cover firmly until the latch ‘clicks’ shut.
Fig. 4-6 Close Station Cover
The Bump / Calibration operation starts up automatically
following the locking of the station cover latch.
Содержание PS200 ABC
Страница 1: ...Personal Surveyor 200 Auto Bump Calibration Station Handbook Gas Measurement Instruments Ltd...
Страница 6: ...iv AUTO BUMP CAL STATION HANDBOOK...
Страница 10: ...viii AUTO BUMP CAL STATION HANDBOOK...
Страница 16: ...1 6 AUTO BUMP CAL STATION HANDBOOK...
Страница 26: ...3 2 AUTO BUMP CAL STATION HANDBOOK Fig 3 2 Windows Vista Fig 3 3 Windows 7...
Страница 54: ...4 14 AUTO BUMP CAL STATION HANDBOOK...
Страница 62: ...5 8 AUTO BUMP CAL STATION HANDBOOK EXAMPLE Fig 5 11...
Страница 69: ...5 15 VIEW PRINT TEST RESULTS Fig 5 17 Bump Results...
Страница 70: ...5 16 AUTO BUMP CAL STATION HANDBOOK Fig 5 18 Calibration Certificate...
Страница 79: ......