B-7
In SPAN mode, the instrument buttons have the functions
shown in Figure B-7.
Single Press
Double Press
Press and
Hold
Button 1
Button 2
Figure B-7
SPAN Mode Button Functions
8a) In SPAN mode, a single press of Button One
will produce small incremental changes to increase
display reading, or a single press of Button Two
will produce small decremental changes to
decrease display reading, until the displayed gas
value corresponds to the concentration of the
calibration gas.
8b) When required reading has been reached, press and
hold Button Two
to exit SPAN mode with
calibration. The display may jump above and below
required reading momentarily as the instrument
performs the calibration.
Note: If for any reason you require to exit SPAN mode
without calibration of the instrument. press and
hold Button One
.
FIELD CALIBRATION
Содержание Gasurveyor 11-500
Страница 1: ...Gasurveyor 11 500 CGI Inert User Handbook...
Страница 5: ...iii DECLARATION OF CONFORMITY...
Страница 6: ...GASURVEYOR 11 500 USER HANDBOOK iv...
Страница 7: ...v REVISION RECORD Date Issue Description Of Change 03 11 06 1 New Handbook...
Страница 8: ...GASURVEYOR 11 500 USER HANDBOOK vi...
Страница 22: ...GASURVEYOR 11 500 USER HANDBOOK 10...
Страница 30: ...GASURVEYOR 11 500 USER HANDBOOK 18...
Страница 38: ...GASURVEYOR 11 500 USER HANDBOOK 26...
Страница 44: ...GASURVEYOR 11 500 USER HANDBOOK 32...
Страница 46: ...GASURVEYOR 11 500 USER HANDBOOK 34...
Страница 50: ...GASURVEYOR 11 500 USER HANDBOOK A 4...
Страница 60: ...GASURVEYOR 11 500 USER HANDBOOK B 10...
Страница 86: ...C 26 GAS MEASUREMENT INSTRUMENTS LTD...
Страница 90: ...GASURVEYOR 11 500 USER HANDBOOK IV...
Страница 91: ......