GASURVEYOR 11-500 USER HANDBOOK
4
Volume Gas, 0 to 100 %
This range displays the total volume of a specific
flammable gas with respect to air. The calibration gas is
shown on the service label and for the purpose of this
handbook is assumed to be methane. Instruments
calibrated for methane in air should only be used for
measuring such mixtures. To change the calibration gas,
e.g. from methane to propane, the instrument must be
recalibrated by suitably trained personnel.
On the Volume Gas range the instrument range is
displayed in the top right corner of the LCD as GAS. The
digital display resolves the signal to 1% GAS with the
analogue bar graph following in steps of 4%. Figure 2.2
shows the Volume Gas display. The detection principle
for the Volume Gas range is thermal conductivity.
Figure 2.2
Volume Gas Range
Oxygen, 0 25%
This range displays the percentage Oxygen content of the
sample. The instrument range is displayed in the top right
corner of the display as % O
2
. From 0 to 21% O
2
the digital
display resolves to 0.1% O
2
. From 21 to 25% O
2
the digital
display resolves to 1% O
2
. The analogue bar graph follows
in 4% steps.
Содержание Gasurveyor 11-500
Страница 1: ...Gasurveyor 11 500 CGI Inert User Handbook...
Страница 5: ...iii DECLARATION OF CONFORMITY...
Страница 6: ...GASURVEYOR 11 500 USER HANDBOOK iv...
Страница 7: ...v REVISION RECORD Date Issue Description Of Change 03 11 06 1 New Handbook...
Страница 8: ...GASURVEYOR 11 500 USER HANDBOOK vi...
Страница 22: ...GASURVEYOR 11 500 USER HANDBOOK 10...
Страница 30: ...GASURVEYOR 11 500 USER HANDBOOK 18...
Страница 38: ...GASURVEYOR 11 500 USER HANDBOOK 26...
Страница 44: ...GASURVEYOR 11 500 USER HANDBOOK 32...
Страница 46: ...GASURVEYOR 11 500 USER HANDBOOK 34...
Страница 50: ...GASURVEYOR 11 500 USER HANDBOOK A 4...
Страница 60: ...GASURVEYOR 11 500 USER HANDBOOK B 10...
Страница 86: ...C 26 GAS MEASUREMENT INSTRUMENTS LTD...
Страница 90: ...GASURVEYOR 11 500 USER HANDBOOK IV...
Страница 91: ......