18
3. Apply Control Solution.
Shake the control solution vial thoroughly before
use. Squeeze out a drop and wipe it off, then
squeeze another drop and place it on the tip of the
vial cap.
Hold the meter to move the absorbent hole of
test strip to touch the drop. Once the confirmation
window fills completely, your meter will begin
counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
2. Press M button to mark this test as a control solution test.
With “
QC
” displayed, the meter will store your test result in memory
under “
QC
”. If you press M button again, the “
QC
” will disappear and
this test is no longer a control solution test.
When doing a control solution test, you have to mark it so that the
test result will NOT mix with the blood glucose TEST RESULTS
stored in the memory. Failure to do so will mix up the blood
glucose test results with the control solution test results in memory.
WARNING