"PROGRAMMING" MENU 

PROGRAMMING MENU 

The Electrostimulator offers the possibility of creating new programs and modifying existing ones.
This makes the PRO models highly flexible and adaptable to all users' requirements.
From the "Programming" entry of the main menu, it is possible to create a new program (when
the word EMPTY appears) and to execute already personalized programs.  These programs can
be modified at any time (see the section  "Modifying an existing program" below).
The programs created with this function are available for all users and cannot be memorized in
the "Last 10 Executed" list nor in the "Favorites" list.  

Create a new program 

From the main menu, move the cursor to "Programming" using the UP and DOWN keys and con-
firm with OK.  Use the UP and DOWN keys to select a number (from 1 to 10) for the program you
wish to create and confirm with OK.  

NOTE:  If the program has already been created, see "Modifying an Existing Program" below.  

To name the new program, use the LEFT and RIGHT keys to select letters and confirm with OK.  To
delete a letter, move the cursor to highlight it and confirm with OK.  

To set up parameters for the new program, use the UP and DOWN keys to select the desired
parameter and confirm with OK.  In the event of an error, press the LEFT key for 2 seconds to return
to the previous display.  The following parameters are set sequentially:

Step 1. Type of stimulation

.  After inserting the name of the new program, use the UP and DOWN

keys to select the desired type of stimulation by choosing among:  EMS, TENS, Microcurrents,
Denervated Muscles.  The type of stimulation available depends on the model of electrostimula-
tor.  Confirm with OK.

Step 2. Number of phases

.  Use the UP and DOWN keys to select the total number of 

phases. Confirm with OK. 

Step 3.  Phase duration.

Use the UP and DOWN keys to select the desired duration (in

minutes)) of the first phase.  Confirm with OK. 

Depending on the type of stimulation selected, you will now be prompted to set several para-
meters regarding the type of current.  Use the UP and DOWN keys to select the desired para-
meter and confirm with OK until finished.  

If the program is composed of multiple phases, after you have set up one phase you will auto-
matically be asked to set up the successive phase.

Modify an existing program 

Within the "Programming" menu, select the number of the program that you wish to modify and
press simultaneously the UP and FUNCTION keys.  At this point it is possible to modify the program,
beginning with its name and passing through steps 1, 2 and 3 above. 

Eliminate a program

Within the "Programming" menu, select the number of the program that you wish to delete and
press simultaneously the DOWN and FUNCTION keys.  It is not possible to recover a program once
it has been eliminated.  

NOTE:  It is not possible to set up mixed, multiphase programs

16

Содержание Genesy 500

Страница 1: ...ELECTROSTIMULATORS Genesy 500 USER MANUAL Rev 03_09 04 0476...

Страница 2: ...33 www globusitalia com info globusitalia com This product has been manufactured according to the technical regulations in force and is certified according to Directive 93 42 EEC for medical devices b...

Страница 3: ...Executed AUTO STIM Automatic Favorites 2 2 Mode simultaneous stimulation 15 PROGRAMMING MENU 16 Create a new program Modify an existing program Eliminate a program ADVANCED MENU 17 INTENSITY OF STIMU...

Страница 4: ...IST INCONTINENCE 25 DIAGNOSTICS 25 Rheobase calculation Chronaxie calculation Slope calculation PROGRAMS LIST DENERVATED 26 ELECTRODE PLACEMENT 27 WARRANTY TERMS 35 FREQUENTLY ASKED QUESTIONS 35 REV 0...

Страница 5: ...resent If any material is missing please contact your retailer immedia tely Please also check the integrity of the stimulator and the supplied electrodes and cables Electrodes The supplied electrodes...

Страница 6: ...bus Electrostimulator be used with caution in persons with capil lary fragility to avoid excessive capillary damage Potential adverse effects Skin irritation may occur in persons with particularly sen...

Страница 7: ...vital functions Store in the supplied carrying case in conditions of 20 85 relative humidity and 5 45 C Recommendations for maintenance and use of electrodes and cables Clean and dry the skin prior t...

Страница 8: ...r the following purposes Pain treatment in general Any type of sequential electrostimulation Aesthetic care Sports training The Electrostimulator may be used with or without the supervision of physici...

Страница 9: ...ates To charge the batteries turn off the electrostimulator and detach the electrodes Attach the supplied battery charger by inserting the plug in the appropriate inlet on the electrostimulator figure...

Страница 10: ...lored cables for EMS and TENS programs Use channels 1 and 3 with the gray cables for the Microcurrents MCR program Use channels 1 and 3 with the colo red cables for the Denervated Muscles program with...

Страница 11: ...s selection 3 during a program Return to the previous phase DOWN DEL Move the selection down During a program simultaneously decrease the intensity of the 4 chan nels TIME During a program displays th...

Страница 12: ...ried out before the message EMPTY appears proceed to the Program List Program selection from the Program List menu Use the UP and DOWN keys to position the cursor on Program List and confirm by press...

Страница 13: ...m press OK Resumption after a pause repeats the last interrupted cycle Stop a program If you need to stop a program before its end switch off the instrument by pressing the OK key for 3 seconds Visual...

Страница 14: ...y execution Registration in memory occurs automatically at the end of any program when the memory is full older programs are automatically deleted To recall a program in memory position the cursor on...

Страница 15: ...ect the program for channels 3 and 4 Use the UP and DOWN keys to select the area of the second program channels 3 and 4 and confirm with OK 6 Use the UP and DOWN keys to select Man or Woman for the se...

Страница 16: ...s display The following parameters are set sequentially Step 1 Type of stimulation After inserting the name of the new program use the UP and DOWN keys to select the desired type of stimulation by cho...

Страница 17: ...up area and allows the duration of back lighting in stand by mode to be varied using the Up and Down keys Contrast FUNCTION The Contrast function present inside the Set up menu allows the level of co...

Страница 18: ...uscle is contracted maximally This is intense treatment that should be performed only only after having carried out various applications at lower intensity In the descriptions of the various treatment...

Страница 19: ...program may offer more than one difficulty level or body area to exercise PROGRAMS LIST PAIN ANTALGIC Endorphinic Tens Contracture Sciatica Lumbago Cervical pain Periathritis S O Epicondylitis Carpal...

Страница 20: ...rou tes of the pain Against every kind of pain It can be used every time you need to relieve pain Warning TENS pro grammes relieve pain but do not eli minate its cause If it persists see the doc tor O...

Страница 21: ...timulator automatically brings the intensity to the correct level This value should not be altered during execution of the program The Microcurrents programs cannot be run in the 2 2 mode with other t...

Страница 22: ...t with metallic items e g tables and chairs Use of iontophoresis programs The duration of iontophoresis programs is automatically set for 10 min Once the program has started it is possible to modify t...

Страница 23: ...equelae of fractu res myalgia osteoarthrosis Thiomucase Lumbago cervical pain cellulitis Sodium Chloride Sclerotic and softener Use with low intensity current Calcium Chloride Increase cal cium conten...

Страница 24: ...aproxen Dioxilase Orudis ketoprofen Dicloreum diclofenac Tantum benzydamine Dievril Xylocaine lidocaine Essaven gel Fastum gel ketoprofen Flectadol lysine acetylsalicylate Fibrase pentosan polysulfate...

Страница 25: ...des it relates to the contractile value of the muscle and to its trophism Chronaxie calculation The instrument automatically selects a value of intensity twice that of the rheobase The ampli tude of t...

Страница 26: ...of the impulse and duration of the rest period Triangular currents These currents reach a maximal value of intensity with a linear ramp that combined with impul ses of rather long duration causes a c...

Страница 27: ...OF ARM BICEPS BRACHII 2 TRICEPS MUSCLE OF ARM TRICEPS BRACHII 3 WRIST AND FINGER EXTENSORS 5 DELTOID MUSCLE 4 WRIST AND FINGER FLEXORS 6 DELTOID MUSCLES ANTERIOR Work position Work position Work posit...

Страница 28: ...STERIOR 8 GREATER AND SMALLER PECTORAL MUSCLES 9 BREAST 12 10 MUSCULUS LATISSIMUS DORSI 11 TRAPEZIUS MUSCLE Work position Work position Work position Work position Work position Work position MUSCULUS...

Страница 29: ...ition Work position Work position Work position Work position Work position MUSCULUS RECTUS ABDOMINIS MUSCULUS LATISSIMUS DORSI AND MUSCULUS ILEOCOSTAL THORACIS MUSCULUS RECTUS ABDOMINIS AND EXTER NAL...

Страница 30: ...EMENT 22 FACE TRAPEZIUS MUSCLES CERVICAL VERTE BRAE 24 PERIARTHRITIS OF SCAPEL OR HUMERUS 21 GLUTEUS MUSCLES 19 QUADRICEPS 20 INTERNAL THIGH Work position Work position Work position Work position Wor...

Страница 31: ...3 09 04 ELECTRODE PLACEMENT 26 EPICONDYLITIS 25 EPICONDYLITIS 27 CERVICOBRACHIAL NEURALGIA 28 BACK 29 SCIATICA 30 KNEE Work position Work position Work position Work position Work position Work positi...

Страница 32: ...REV 03 09 04 32 Work position Work position Work position Work position Work position Work position 31 32 33 34 35 36 ELECTRODE PLACEMENT for microcurrents...

Страница 33: ...REV 03 09 04 ELECTRODE PLACEMENT for microcurrents Work position Work position Work position Work position Work position Work position 33 37 38 39 40 41 42...

Страница 34: ...ELECTRODE PLACEMENT with accessories Work position Work position Work position Work position Work position Work position 43 44 45 46 47 48 REV 03 09 04 34...

Страница 35: ...ram check that the cables are connected and that the electrodes are positioned on the area to be treated In addition be sure that the electrodes are not worn because this could inhibit their ability t...

Страница 36: ...GLOBUS ITALIA SRL Via Vittorio Veneto 52 31013 TV Tel 0438 7933 Fax 0438 793363 www globusitalia com info globusitalia com...

Отзывы: