VI.

 

Computer Operation 

 

Main Display 

 
In the main display, the 4 buttons are 
labeled as shown.   
 
The MODE button toggles the  
upper display between average  
velocity minimum, maximum  
and timer.  The label next to  
this display changes to correctly  
indicate which parameter is shown. 
 
The MENU button is used to 
view stored data, delete data, 
set the flow units to FT/S or M/S, 
and change the calibration setting. 
Press the MENU button for 2 seconds 
to enter the setup menu.  In menu mode, the labeling of the 4 buttons 
on the LCD display may change to indicate different functions. 
 
The SAVE button is used to save the average, minimum and 
maximum velocity data to memory. 
 
The RESET button causes the average, minimum and maximum data 
to be set to the current instantaneous velocity shown in the large 
lower display.  Averaging and storage of minimum and maximum 
values begins at that time at the rate of once per second.   
 
Note that the propeller should already be in position to begin taking 
measurements when the RESET button is pressed; if it isn’t, the 
button can be pressed again.  When the RESET button is pressed 
while the propeller is not turning, the avg/min/max parameters are all 
set to zero.  This will cause the minimum velocity to always read zero 
and the average velocity will include all the zero readings since the 

Содержание FP111

Страница 1: ...r Instrumentation Visit Our Complete Online Catalog www globalw com Call 800 876 1172 7 30 AM to 4 PM Pacific Time International US 916 638 3429 FAX 916 638 3270 E mail globalw globalw com 11390 Amalg...

Страница 2: ...e a valuable asset for your application Should you require assistance our technical staff will be happy to help Table of Contents I Checklist Page 3 II Inspection 3 III Description 3 IV General Instru...

Страница 3: ...list to ensure that you have received everything needed to operate the Flow Probe III Description The Global Water Flow Probe is a rugged and highly accurate water velocity instrument for measuring fl...

Страница 4: ...usly stored data The flow probe computer is self contained and sealed It has an internal lithium battery that should last approximately 5 years depending on use The computer floats and is shock and wa...

Страница 5: ...updated once per second Pressing RESET before beginning a measurement is required for best accuracy e To determine flow from the velocity data measure or calculate the cross sectional area of your fl...

Страница 6: ...the flow through a series of flow surges c The water is not always moving in the same direction Even the smoothest water flow contains turbulence that causes the water to move in different directions...

Страница 7: ...ge minimum maximum velocities and a running average is started As long as the probe remains in the flow the averaging continues To stop averaging press the SAVE button in Save Data mode the averaging...

Страница 8: ...ow for the subsection Q V x A Once the flow of each subsection is obtained add all of the subsection flows to obtain the total stream flow c For the USGS 6 tens method the Flow Probe is placed at the...

Страница 9: ...to indicate different functions The SAVE button is used to save the average minimum and maximum velocity data to memory The RESET button causes the average minimum and maximum data to be set to the cu...

Страница 10: ...parameters to the memory location shown in the upper display and the computer returns to the main display mode automatically resetting the average minimum and maximum values The memory location is au...

Страница 11: ...enu function View Data Delete All Set Units Set Cal View Data From the View Data menu screen press SET to review recorded data The memory location is shown in the upper display and the data is shown i...

Страница 12: ...nd return to the menu screen Press SET to delete all data from memory Once deleted the data cannot be recovered Set Flow Units From the Set Units menu display Press SET to change the Flow Units The lo...

Страница 13: ...400 The calibration factor scales rotational speed to velocity in FT S only If the flow units are set to M S meters per second the calibration factor does not need to be changed Conversion between FT...

Страница 14: ...carbonate housing with polyester overlay Power Internal Lithium Approx 5 year life Non Replaceable Operating Temperature 20 to 70 C 4 to 158 F Non Freezing Storage Temperature 30 to 80 C 22 to 176 F V...

Страница 15: ...cycled at no charge c Cleaning Make sure the Turbo Prop turns freely before and after your measurements Blow on the prop in the direction of flow The prop should turn freely If not rinse the probe in...

Страница 16: ...te supplied with the flow probe It can also be obtained by contacting the Technical Support department at Global Water be ready to supply the serial number for the flow probe e If the display is blank...

Страница 17: ...e incurred during transit X Warranty a Global Water Instrumentation Inc warrants that its products are free from defects in material and workmanship under normal use and service for a period of one ye...

Страница 18: ...16 0 0811 0 66 0 5499 GPM x 1440 Gallons day GPD 0 17 0 0885 0 67 0 5594 0 18 0 0961 0 68 0 5687 0 19 0 1039 0 69 0 5780 0 20 0 1118 0 70 0 5872 0 21 0 1199 0 71 0 5964 0 22 0 1281 0 72 0 6054 0 23 0...

Страница 19: ...ng standard EN 61326 1 2006 Battery and Accumulators Directive 2006 66 EC Requirements regarding the proper handling and disposal of batteries and accumulators Due to the rugged conditions this produc...

Отзывы: