Operation 

 

Operation of the control is regulated by the operator and does not utilize a sensing element. 

After a desired temperature has been determined on the initial run, a simple resetting of the dial 
offers control for repeated operations. 

1.

 

Turn control knob to OFF position. 

2.

 

Plug device into the 3-wire outlet of the control. 

3.

 

Insert plug of the control cord into 3-wire wall or bench receptacle. 

4.

 

Adjust knob to desired setting. The pilot (cycle) light is connected across the output side of 
the control and indicates power to load. 

PL312 & PL3122 

 

PL324 

Dial Setting 

(Réglage du 

cadrant) 

Time (Temps) in Seconds 

 

Dial Setting 
(Réglage du 

cadrant) 

Time (Temps) in Seconds 

On 

Off 

Total 

Cycle 

% ACTIVE 

(Pourcentage de 

passage du courant) 

 

On 

Off 

Total 

Cycle 

% ACTIVE 

(Pourcentage de 

passage du courant) 

 

LO 

4.00 

50.00 

54.00 

7.41% 

 

LO 

4.00 

53.00 

57.00 

7.02% 

4.00 

50.00 

54.00 

7.41% 

 

4.00 

48.00 

52.00 

7.69% 

4.20 

31.00 

35.20 

11.93% 

 

4.10 

23.00 

27.10 

15.13% 

4.60 

20.10 

24.70 

18.62% 

 

5.20 

13.00 

18.20 

28.57% 

5.20 

14.60 

19.80 

26.26% 

 

5.50 

11.00 

16.50 

33.33% 

6.00 

11.90 

17.90 

33.52% 

 

7.00 

9.00 

16.00 

43.75% 

6.40 

10.40 

16.80 

38.10% 

 

9.00 

6.50 

15.50 

58.06% 

7.50 

9.00 

16.50 

45.45% 

 

13.00 

5.50 

18.50 

70.27% 

9.80 

7.50 

17.30 

56.65% 

 

17.00 

5.00 

22.00 

77.27% 

Does not 

Go off 

Does not 

Go off 

Does not 

Go off 

100% 

 

Does not 

Go off 

Does not 

Go off 

Does not 

Go off 

100% 

HI 

Does not 

Go off 

Does not 

Go off 

Does not 

Go off 

100% 

 

HI 

Does not 

Go off 

Does not 

Go off 

Does not 

Go off 

100% 

The control should not be exposed to ambient temperature above 125

O

 F (52

 C) 

Ces régulateurs ne doivant pas être exposés à des températures de plus de 125

 F (52

 C) 

 

Description et mode d’emploi des régulateurs de puissance Minitrol et Minitwin de Glas-Col 

 

Numéros de catalogue 

Puissance 

PL312  

 

 

1440 watts, 12A, 120V 

PL3122 

 

 

1800 watts, 15A, 120V 

PL324  

 

 

2400 watts, 10A, 240V 

 
 

Les régulateurs de puissance Minitrol et Minitwin, à commande manuelle, fonctionnent en 

coupant et rétablissant le courant alternativement, c’est-à-dire en augmentant ou diminuant le temps 
de passage du courant, à plein voltage, de 5-1/2% à 100% selon la position de l’index sur le cadran. 
La position OFF (hors circuit) coupe le courant des deux côtes. Ces régulateurs peuvent être utilisés 
jusqu’à leur puissance nominale pour toute charge à résistance non-inductive; on peut se servir de 
fusibles de moindre ampérage pour protéger des appareils de très faible puissance. 
 
 

Ces régulateurs sont destinés à régler l’allure des chauffe-ballons, rubans et cordons 

chauffants de Glas-Col, mais peuvent servir aussi pour régler des fours et des étuves de faibles 
dimensions, des plaques chauffantes, moules et autres appareils. 
 