USER’S MANUAL
TC15
9. CALIBRATION seTTINgs
1. When the user is in the normal mode of weighing, he must press the
and
,
the message
ECF – 1
is going to appear
on the display.
2. Press the
or
to choose the desired function:
ECF-1, ECF-2 o ECF-3
* ECF-1 CALIBrATION OF ZErO + wEIGhT
Press the key
to access.
Press the key
to confirm the calibration of 0.
Press the key
Numerical keyboards
to introduce the value of the weight of calibration.
Place the weight of calibration on the platform and press the key
to do the calibration once the reading is
steady.
* ECF-2 CALIBrATION OF ZErO
Press the key
to accesss.
Press the key
to “0” calibration.
* ECF-3 CALIBrATION OF wEIGhT (SPAN)
Press the Key
the display will show the value of the weight of calibration.
Press the
Numerical keyboards
, to modify the value of the weight of calibration.
Place the weight of calibration on the platform and press the key
to do the calibration once the reading is
stable..
10. TeChNICAL PARAMeTeRs
DO NOt MODIfY tHE tECHNICAL PARAMEtERS If It IS NOt StRICtLY NEEDED. A BAD CONfIgURAtION Of tHIS SECtION CAN
CAUSE A WRONg fUNCtIONINg Of tHE SCALE.
ENTRANCE AND EXIT TECHNICAL PARAmETERS
dISPLAy
dESCrIPTION
P 0000
With the power off, Press hold on
until the display shows:
P 0000
0020
Enter the value 0020 with
Numerical keyboards
and press
to access to menu.
LF1
the display shows [
LF1
]
LF 2
Press on the key
or
, until the display show [ Lf desired]
Press
to acess the Lf desired.