EN
USER’S MANUAL
ABS
76
5.2 CALIBRATION / ADJUSTMENT
REQUEST
The machine only can be Calibration / Adjustment in the following cases:
• Nothing on weighing pan, was tare, weighing signal stable
Machine will have ERROR SIGNAL if does not have above premise.
If all premise available, the machine will display required weight mass value for calibration.
INSTANCE
EXTERNAL SPAN CALIBRATION INSTANCE
KEY (ORDER)
STEP EXPLANATION
LCD SCREEN DISPLAY
1. Machine tare the weight
0.0000 g
Press and Hold [CAL] Key
2. Display span calibration signal Flash the required weight mass value:
100g
100.0000 g
3. Put required weight mass on pan. After 5 seconds, display the weight
of it
100.0000 g
4. Take the weight mass away (Span calibration finished)
0.0000 g
EXTERNAL LINEARITY CALIBRATION INSTANCE
KEY (ORDER)
STEP EXPLANATION
LCD SCREEN DISPLAY
Short Press [TARE] Key
1. Machine tare
0.0000 g
Press and Hold [CAL] Key for
5 seconds
2. Display Span CAL signal Flash 100g signal
100.0000 g
Release [CAL] Key,
Press and Hold [MENU] Key
for 5 seconds
3. Display Linearity CAL signal Flash 100g signal
100.0000 g
4. Put required weight mass on pan Display 100g after 5 seconds
100.0000 g
5. Take weight mass away Linearity calibration step Flash 50g signal
50.0000 g
6. Put required weight mass on pan Display 50g after 5 seconds
50.0000 g
7. Take weight mass away (Linearity calibration finished)
0.0000 g
NOTE:
The grey color words explain the signal’s meaning which flash on the window.