GB
22
If the overfill security system it’s activated, don’t proceede with the liquid aspiration. If the overfill security system doesn’t work
there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, in this case return the
device to
GIMA
technical service.
GIMA S.p.A.
will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
GIMA
TECHNICAL SERVICE.
GIMA
DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
INSTRUCTION FOR USE
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the LCD screen, the
suction tank and the antibacterial filter.
WARNING:
For proper use, place the aspirator on a flat, stable surface
in order to have the full volume of use of the jar and
better efficiency of the overflow device.
Operation using with AC/DC switching adapter
Connect the short silicon tube
,
with antibacterial filter, to the suction connector (VACUUM). The other tube, with one end
connected to the antibacterial filter with the other end to jar’s lid connector where has been fixed the red float. When the
90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on
the jar) to the avoid liquid penetration inside the device.
Connect the long silicon tube to the other jar's lid connector (PATIENT PORT);
Connect the switching adapter to the device with the appropriate connector and insert the power cable plug to the power
socket. To start the treatment press the ON/OFF button and check its green back-light. When the device is switched on the
LCD screen activates displaying a bar-graph, graduated scale, and negative pressure value, which must be set by the operator;
The negative pressure is set with keys - and +, which are found below the LCD screen: Press one of the two keys with the
motor on. The level of the negative pressure will flash at the centre of the display.
Release the key and, after 3 seconds, the value will be operating (fixed digit on the display) and saved. The desired value
remains fixed until the operator varies the negative pressure, whilst the bar-graph moves along the semi-circumference
stopping in correspondence (graph) of the value set.
The maximum negative pressure value that can be set is: -75kPa (-0.75 bar)
The minimum negative pressure value that can be set is: -10kPa (-0.10 bar)
To suspend and / or terminate the treatment press the ON / OFF button and and pull the plug out from the power socket.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of
the functionally vacuum) then rescrew the lid on the jar correctly.
To extract the accessories and start with cleaning.
The battery is fully charged in about 70 minutes with continuous operation.
WARNING:
Before using the device, check the battery power status. Before each use proceed with charging the
battery. To maintain the device in good conditions, recharge the battery every 3 months (when not in use).
WARNING:
Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
“VACUUM”. A wrong connection causes immediate destruction in case of contact with sucked liquids.
WARNING:
The power supply cable plug is the element of separation from the electrical mains system: even if the units equipped
with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to allow a
further method of disconnection from the mains supply system.