background image

24

Guidance and manufacturer’s declaration – Immunity Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below.

The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it’s used in such an 

environment.
Immunity Test

Level indicated by the

EN 60601-1-2

Compliance 

Level

Electromagnetic environment - guidance 

Conducted 

Immunity

EN 61000-4-6

Radiated 

Immunity

EN 61000-4-3

3Vrms 150kHz 

to 80Mhz

(for non life-supporting

devices)

3V/m 80MHz 

to 2.7GHz

(for non life-supporting

devices)

V1 = 3 V rms

E1 = 10 V / m

The  portable  and  mobile  RF  communication  devices, 

including cables, must not be used closer to the SU-

PERVEGA BATTERY SUCTION UNIT device, than the 

separation distance calculated by the equation applica-

ble to the transmitter frequency.

Recommended separation distance

 

from 80 MHz to 800MHz
 

from 800 MHz to 2.7 GHz

Where 

P

 is the maximum nominal output voltage of the

transmitter in Watt (W) depending on the manufactur-

er of the transmitter and the recommended separation 

distance in metres (m). The intensity of the field from 

the  fixed  RF  transmitters,  as  determined  by  an  elec-

tro-magnetic study of the site

a)

, could be lower than the 

level of conformity of each frequency interval 

b)

It is possible to check for interference in 

proximity to devices identified by the fol-

lowing symbol:

Note 1:

 

At 80 MHz and 800 MHz the interval with the highest frequency is applied

Note 2:

 

These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the

absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terres-

trial mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and 

accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnet-

ic study of the site should be considered. If the field intensity measured in the place where the device will be used surpass-

es the above mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnor-

mal performance arises, additional measures such as changing the device’s direction or positioning may be necessary.

b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.

3,5

 V

1

√P

d

=

12

 E

1

√P

d=

 

23

 E

1

√P

d=

ENGLISH

Содержание SUPERVEGA 28193

Страница 1: ...EUR SUPERVEGA BATTERY SUR CHARIOT ABSAUGER SUPERVEGA BATTERY AUF TROLLEY ASPIRADOR SUPERVEGA BATTERY SOBRE CARRO M28193 M Rev 1 01 21 PROFESSIONAL MEDICAL PRODUCTS 0476 Gima S p A Via Marconi 1 20060...

Страница 2: ...inlets on the back aren t obstructed To avoid incidents do not place the aspirator on unstable surfaces which may cause it to accidentally fall and lead to a malfunction and or breakage Should there...

Страница 3: ...in chest fluids The device must not be used for suction of explosive corrosive or easily flammable liquids SUPERVEGA BATTERY SUCTION UNIT is not suitable for MRI Do not introduce the device in MRI env...

Страница 4: ...use The filter is not designed to be decontaminated disassembled and or sterilised If the patient s pathology is known and or no risk of indirect contamination exists it is advisable to replace the f...

Страница 5: ...rations according to the following procedure Place the overflow valve into its seat in the cover under VACUUM connector Insert floating valve keeping the o ring towards the opening of the cage Place t...

Страница 6: ...ttery pay special attention to the polarity of the same component The polarities are indicated directly on the battery If the overfill security system it s activated don t proceede with the liquid asp...

Страница 7: ...se must be used in vertical mode to prevent the action of the backflow valve If this protection is triggered turn the device off and disconnect the pipe connected to the vacuum jar indicated with the...

Страница 8: ...able to charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx 240 minutes with the main switch to position 0 TAB I INDICATOR LIGHTS...

Страница 9: ...e float valve intervenes signalling the device is too full the suction source must be disconnected within no more than 5minutes RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES This section...

Страница 10: ...nt Immunity Test Level indicated by the EN 60601 1 2 Compliance Level Electromagnetic environment guidance Electrostatic discharge ESD EN 61000 4 2 8kV on contact 15kV in air The device doesn t change...

Страница 11: ...mined by an elec tro magnetic study of the sitea could be lower than the level of conformity of each frequency interval b It is possible to check for interference in proximity to devices identified by...

Страница 12: ...2 12 23 For transmitters with a maximum nominal output power not shown above the recommended separation distance in metres m can be calculated using the equation applicable to the transmitter frequenc...

Страница 13: ...rectly you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery The recycling of the...

Отзывы: