Table 1: Guidance and manufacturer’s declaration - electromagnetic
emissions
The device is intended for use in the electromagnetic environment specified below.
The user of this device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1
This device uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all
establishments. It uses internal power supply
and has no connection with power supply
network.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
N/A
Table 2: Guidance and manufacturer’s declaration - electromagnetic
immunity
The device is intended for use in the electromagnetic environment specified below.
The user of this device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for
power supply lines
±1 kV for
input/output lines
N/A
N/A